07.08.2019 | Hepatobiliary-Pancreas

Radiomic feature reproducibility in contrast-enhanced CT of the pancreas is affected by variabilities in scan parameters and manual segmentation

verfasst von: Rikiya Yamashita, Thomas Perrin, Jayasree Chakraborty, Joanne F. Chou, Natally Horvat, Maura A. Koszalka, Abhishek Midya, Mithat Gonen, Peter Allen, William R. Jarnagin, Amber L. Simpson, Richard K. G. Do

Erschienen in: European Radiology | Ausgabe 1/2020

Einloggen, um Zugang zu erhalten

Abstract

Objectives

This study aims to measure the reproducibility of radiomic features in pancreatic parenchyma and ductal adenocarcinomas (PDAC) in patients who underwent consecutive contrast-enhanced computed tomography (CECT) scans.

Methods

In this IRB-approved and HIPAA-compliant retrospective study, 37 pairs of scans from 37 unique patients who underwent CECTs within a 2-week interval were included in the analysis of the reproducibility of features derived from pancreatic parenchyma, and a subset of 18 pairs of scans were further analyzed for the reproducibility of features derived from PDAC. In each patient, pancreatic parenchyma and pancreatic tumor (when present) were manually segmented by two radiologists independently. A total of 266 radiomic features were extracted from the pancreatic parenchyma and tumor region and also the volume and diameter of the tumor. The concordance correlation coefficient (CCC) was calculated to assess feature reproducibility for each patient in three scenarios: (1) different radiologists, same CECT; (2) same radiologist, different CECTs; and (3) different radiologists, different CECTs.

Results

Among pancreatic parenchyma-derived features, using a threshold of CCC > 0.90, 58/266 (21.8%) and 48/266 (18.1%) features met the threshold for scenario 1, 14/266 (5.3%) and 15/266 (5.6%) for scenario 2, and 14/266 (5.3%) and 10/266 (3.8%) for scenario 3. Among pancreatic tumor-derived features, 11/268 (4.1%) and 17/268 (6.3%) features met the threshold for scenario 1, 1/268 (0.4%) and 5/268 (1.9%) features met the threshold for scenario 2, and no features for scenario 3 met the threshold, respectively.

Conclusions

Variations between CECT scans affected radiomic feature reproducibility to a greater extent than variation in segmentation. A smaller number of pancreatic tumor-derived radiomic features were reproducible compared with pancreatic parenchyma-derived radiomic features under the same conditions.

Key Points

• For pancreatic-derived radiomic features from contrast-enhanced CT (CECT), fewer than 25% are reproducible (with a threshold of CCC < 0.9) in a clinical heterogeneous dataset.
• Variations between CECT scans affected the number of reproducible radiomic features to a greater extent than variations in radiologist segmentation.
• A smaller number of pancreatic tumor-derived radiomic features were reproducible compared with pancreatic parenchyma-derived radiomic features under the same conditions.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
9.
26.
Zurück zum Zitat Pietikäinen M, Hadid A, Zhao G, Ahonen T (2011) Local binary patterns for still images. In: Computer vision using local binary patterns. Springer, London, pp 13–47CrossRef Pietikäinen M, Hadid A, Zhao G, Ahonen T (2011) Local binary patterns for still images. In: Computer vision using local binary patterns. Springer, London, pp 13–47CrossRef
27.
Zurück zum Zitat Mehta R, Egiazarian K (2013) Rotated local binary pattern (RLBP): rotation invariant texture descriptor. In 2nd International Conference on Pattern Recognition Applications and Methods, ICPRAM 2013, Barcelona, Spain, 15.-18.2.2013, pp 497-502. (International Conference on Pattern Recognition Applications and Methods). Institute of Electrical and Electronics Engineers IEEE Mehta R, Egiazarian K (2013) Rotated local binary pattern (RLBP): rotation invariant texture descriptor. In 2nd International Conference on Pattern Recognition Applications and Methods, ICPRAM 2013, Barcelona, Spain, 15.-18.2.2013, pp 497-502. (International Conference on Pattern Recognition Applications and Methods). Institute of Electrical and Electronics Engineers IEEE
31.
Zurück zum Zitat Chakraborty J, Rangayyan RM, Banik S, et al (2012) Detection of architectural distortion in prior mammograms using statistical measures of orientation of texture. Proc. SPIE 8315, Medical Imaging 2012: Computer-Aided Diagnosis, 831521 (23 February 2012). https://doi.org/10.1117/12.910937 Chakraborty J, Rangayyan RM, Banik S, et al (2012) Detection of architectural distortion in prior mammograms using statistical measures of orientation of texture. Proc. SPIE 8315, Medical Imaging 2012: Computer-Aided Diagnosis, 831521 (23 February 2012). https://​doi.​org/​10.​1117/​12.​910937
34.
39.
Zurück zum Zitat Zwanenburg A, Leger S, Vallières M, et al (2016) Image biomarker standardisation initiative. arXiv:1612.07003v7 [cs.CV] Zwanenburg A, Leger S, Vallières M, et al (2016) Image biomarker standardisation initiative. arXiv:1612.07003v7 [cs.CV]
Metadaten
Titel
Radiomic feature reproducibility in contrast-enhanced CT of the pancreas is affected by variabilities in scan parameters and manual segmentation
verfasst von
Rikiya Yamashita
Thomas Perrin
Jayasree Chakraborty
Joanne F. Chou
Natally Horvat
Maura A. Koszalka
Abhishek Midya
Mithat Gonen
Peter Allen
William R. Jarnagin
Amber L. Simpson
Richard K. G. Do
Publikationsdatum
07.08.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 1/2020
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-019-06381-8

Weitere Artikel der Ausgabe 1/2020

European Radiology 1/2020 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

„Nur wer sich gut aufgehoben fühlt, kann auch für Patientensicherheit sorgen“

13.04.2024 Klinik aktuell Kongressbericht

Die Teilnehmer eines Forums beim DGIM-Kongress waren sich einig: Fehler in der Medizin sind häufig in ungeeigneten Prozessen und mangelnder Kommunikation begründet. Gespräche mit Patienten und im Team können helfen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.