Retrospective fractional dose reduction in Tc-99m cardiac perfusion SPECT/CT patients: A human and model observer study

verfasst von: P. Hendrik Pretorius, PhD, Albert Juan Ramon, PhD, Michael A. King, PhD, Arda Konik, PhD, Seth T. Dahlberg, MD, Mathew W. Parker, MD, Naomi F. Botkin, MD, Karen L. Johnson, NMT, Yongyi Yang, PhD, Miles N. Wernick, PhD

Erschienen in: Journal of Nuclear Cardiology | Ausgabe 2/2021

Einloggen, um Zugang zu erhalten

Abstract

Background

In the ongoing efforts to reduce cardiac perfusion dose (injected radioactivity) for conventional SPECT/CT systems, we performed a human observer study to confirm our clinical model observer findings that iterative reconstruction employing OSEM (ordered-subset expectation-maximization) at 25% of the full dose (quarter-dose) has a similar performance for detection of hybrid cardiac perfusion defects as FBP at full dose.

Methods

One hundred and sixty-six patients, who underwent routine rest-stress Tc-99m sestamibi cardiac perfusion SPECT/CT imaging and clinically read as normally perfused, were included in the study. Ground truth was established by the normal read and the insertion of hybrid defects. In addition to the reconstruction of the 25% of full-dose data using OSEM with attenuation (AC), scatter (SC), and spatial resolution correction (RC), FBP and OSEM (with AC, SC, and RC) both at full dose (100%) were done. Both human observer and clinical model observer confidence scores were obtained to generate receiver operating characteristics (ROC) curves in a task-based image quality assessment.

Results

Average human observer AUC (area under the ROC curve) values of 0.725, 0.876, and 0.890 were obtained for FBP at full dose, OSEM at 25% of full dose, and OSEM at full dose, respectively. Both OSEM strategies were significantly better than FBP with P values of 0.003 and 0.01 respectively, while no significant difference was recorded between OSEM methods (P = 0.48). The clinical model observer results were 0.791, 0.822, and 0.879, respectively, for the same patient cases and processing strategies used in the human observer study.

Conclusions

Cardiac perfusion SPECT/CT using OSEM reconstruction at 25% of full dose has AUCs larger than FBP and closer to those of full-dose OSEM when read by human observers, potentially replacing the higher dose studies during clinical reading.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat IAEA. Nuclear cardiology: Its role in cost effective care. IAEA human health series no 18, Vienna, Austria 2012. IAEA. Nuclear cardiology: Its role in cost effective care. IAEA human health series no 18, Vienna, Austria 2012.
2.
Zurück zum Zitat Berrington de Gonzalez A, Kim K, Smith-Bindman R, McAreavey D. Myocrdial perfuion scans: Projected population cancer risks from current levels of use in the United States. Circulation 2010;122:2403-10. Berrington de Gonzalez A, Kim K, Smith-Bindman R, McAreavey D. Myocrdial perfuion scans: Projected population cancer risks from current levels of use in the United States. Circulation 2010;122:2403-10.
3.
Zurück zum Zitat Siegel JA, Pennington CW, Sacks B. Subjecting radiologic imaging to the linear no-threshold hypothesis: A non sequitir of non-trivial proportion. J Nucl Med 2017;58:1-6.CrossRef Siegel JA, Pennington CW, Sacks B. Subjecting radiologic imaging to the linear no-threshold hypothesis: A non sequitir of non-trivial proportion. J Nucl Med 2017;58:1-6.CrossRef
4.
Zurück zum Zitat Duncan JR, Lieber MR, Adachi N, Wahl RL. Radiation dose does matter: Mechanistic insights into DNA damage and repair support the linear no-threshold model of low-dose radiation health risks. J Nucl Med 2018;59:1014-6.CrossRef Duncan JR, Lieber MR, Adachi N, Wahl RL. Radiation dose does matter: Mechanistic insights into DNA damage and repair support the linear no-threshold model of low-dose radiation health risks. J Nucl Med 2018;59:1014-6.CrossRef
5.
Zurück zum Zitat Einstein AJ, Pascual TNB, Mercuri M, Karthikeyan G, Vitola JV, Mahmarian JJ, et al. Current worldwide nuclear cardiology practices and radiation exposure: Results from the 65 country IAEA nuclear cardiology protocols cross-sectional study (INCAPS). Eur Heart J 2015;36:1689-96.CrossRef Einstein AJ, Pascual TNB, Mercuri M, Karthikeyan G, Vitola JV, Mahmarian JJ, et al. Current worldwide nuclear cardiology practices and radiation exposure: Results from the 65 country IAEA nuclear cardiology protocols cross-sectional study (INCAPS). Eur Heart J 2015;36:1689-96.CrossRef
6.
Zurück zum Zitat Abbott BG, Case JA, Dorbala S, Einstein AJ, Galt JR, Pagnanelli R, et al. Contemporary cardiac SPECT imaging-innovations and best practices: An information statement from the American Society of Nuclear Cardiology. J Nucl Cardiol 2018;25:1847-60.CrossRef Abbott BG, Case JA, Dorbala S, Einstein AJ, Galt JR, Pagnanelli R, et al. Contemporary cardiac SPECT imaging-innovations and best practices: An information statement from the American Society of Nuclear Cardiology. J Nucl Cardiol 2018;25:1847-60.CrossRef
7.
Zurück zum Zitat Bateman TM, Heller GV, McGhie AI, Courter SA, Golub RA, Case JA, et al. Multicenter investigation comparing a highly efficient half-time stress-only attenuation correction approach against standard rest-stress Tc-99m SPECT imaging. J Nucl Cardiol 2009;16:726-35.CrossRef Bateman TM, Heller GV, McGhie AI, Courter SA, Golub RA, Case JA, et al. Multicenter investigation comparing a highly efficient half-time stress-only attenuation correction approach against standard rest-stress Tc-99m SPECT imaging. J Nucl Cardiol 2009;16:726-35.CrossRef
8.
Zurück zum Zitat DePuey EG, Bommireddipalli S, Clark J, Thompson L, Srour Y. Wide beam reconstruction “quarter-time” gated myocardial perfusion SPECT functional imaging: A comparison to “full-time” ordered subset expectation maximum. J Nucl Cardiol 2009;16:736-52.CrossRef DePuey EG, Bommireddipalli S, Clark J, Thompson L, Srour Y. Wide beam reconstruction “quarter-time” gated myocardial perfusion SPECT functional imaging: A comparison to “full-time” ordered subset expectation maximum. J Nucl Cardiol 2009;16:736-52.CrossRef
9.
Zurück zum Zitat DePuey EG, Bommireddipalli S, Clark J, Leykekhman A, Thompson LB, Friedman M. A comparison of the image quality of full-time myocardial perfusion SPECT vs wide beam reconstruction half-time and half-dose SPECT. J Nucl Cardiol 2011;18:273-80.CrossRef DePuey EG, Bommireddipalli S, Clark J, Leykekhman A, Thompson LB, Friedman M. A comparison of the image quality of full-time myocardial perfusion SPECT vs wide beam reconstruction half-time and half-dose SPECT. J Nucl Cardiol 2011;18:273-80.CrossRef
10.
Zurück zum Zitat Enevoldsen LH, Menashi CA, Andersen UB, Jensen LT, Henriksen OM. Effects of acquisition time and reconstruction algorithm on image quality, quantitative parameters, and clinical interpretation of myocardial perfusion imaging. J Nucl Cardiol 2013;20:1086-92.CrossRef Enevoldsen LH, Menashi CA, Andersen UB, Jensen LT, Henriksen OM. Effects of acquisition time and reconstruction algorithm on image quality, quantitative parameters, and clinical interpretation of myocardial perfusion imaging. J Nucl Cardiol 2013;20:1086-92.CrossRef
11.
Zurück zum Zitat Gutstein A, Navzorov R, Solodky A, Mats I, Kornowski R, Zafrir N. Angiographic correlation of myocardial perfusion imaging with half the radiation dose using ordered-subset expectation maximization with resolution recovery software. J Nucl Cardiol 2013;20:539-44.CrossRef Gutstein A, Navzorov R, Solodky A, Mats I, Kornowski R, Zafrir N. Angiographic correlation of myocardial perfusion imaging with half the radiation dose using ordered-subset expectation maximization with resolution recovery software. J Nucl Cardiol 2013;20:539-44.CrossRef
12.
Zurück zum Zitat Zafrir N, Bental T, Solodky A, Ben-Shlomo A, Mats I, Hassid Y, et al. Feasibility of myocardial perfusion imaging with half the radiation dose in obese patients using ordered-subset expectation maximization with resolution recovery software. J Nucl Cardiol 2013;20:111-9.CrossRef Zafrir N, Bental T, Solodky A, Ben-Shlomo A, Mats I, Hassid Y, et al. Feasibility of myocardial perfusion imaging with half the radiation dose in obese patients using ordered-subset expectation maximization with resolution recovery software. J Nucl Cardiol 2013;20:111-9.CrossRef
13.
Zurück zum Zitat Oddstig J, Hedeer F, Jogi J, Carlsson M, Hindorf C, Engblom H. Reduced administered activity, reduced acquisition time, and preserved image quality for the new CZT camera. J Nucl Cardiol 2013;20:38-44.CrossRef Oddstig J, Hedeer F, Jogi J, Carlsson M, Hindorf C, Engblom H. Reduced administered activity, reduced acquisition time, and preserved image quality for the new CZT camera. J Nucl Cardiol 2013;20:38-44.CrossRef
14.
Zurück zum Zitat Patil HR, Bateman TM, McGhie AI, Burgett EV, Courter SA, Case JA, et al. Diagnostic accuracy of high-resolution attenuation-corrected Anger-camera SPECT in detection of coronary artery disease. J Nucl Cardiol 2014;21:127-34.CrossRef Patil HR, Bateman TM, McGhie AI, Burgett EV, Courter SA, Case JA, et al. Diagnostic accuracy of high-resolution attenuation-corrected Anger-camera SPECT in detection of coronary artery disease. J Nucl Cardiol 2014;21:127-34.CrossRef
15.
Zurück zum Zitat Oddstig J, Hindorf C, Hedeer F, Jogi J, Arheden H, Hansson MJ, et al. The radiation dose to overweighted patients undergoing myocardial perfusion SPECT can be significantly reduced: Validation of a linear weight-adjusted activity administration protocol. J Nucl Cardiol 2017;24:1912-21.CrossRef Oddstig J, Hindorf C, Hedeer F, Jogi J, Arheden H, Hansson MJ, et al. The radiation dose to overweighted patients undergoing myocardial perfusion SPECT can be significantly reduced: Validation of a linear weight-adjusted activity administration protocol. J Nucl Cardiol 2017;24:1912-21.CrossRef
16.
Zurück zum Zitat van Dijk JD, Lager PL, Ottervanger JP, de Boer J, Oostdijk AHJ, Engbers EM, et al. Development and validation of patient-tailored dose regime in myocardial perfusion imaging using conventional SPECT. J Nucl Cardiol 2016;23:134-42.CrossRef van Dijk JD, Lager PL, Ottervanger JP, de Boer J, Oostdijk AHJ, Engbers EM, et al. Development and validation of patient-tailored dose regime in myocardial perfusion imaging using conventional SPECT. J Nucl Cardiol 2016;23:134-42.CrossRef
17.
Zurück zum Zitat Juan Ramon A, Yang Y, Pretorius PH, Slomka PJ, Johnson KL, King MA, et al. Investigation of dose reduction in cardiac perfusion SPECT via optimization and choice of the image reconstruction strategy. J Nucl Cardiol 2018;25:2217-8.CrossRef Juan Ramon A, Yang Y, Pretorius PH, Slomka PJ, Johnson KL, King MA, et al. Investigation of dose reduction in cardiac perfusion SPECT via optimization and choice of the image reconstruction strategy. J Nucl Cardiol 2018;25:2217-8.CrossRef
18.
Zurück zum Zitat Chang SM, Nabi F, Xu J, Raza U, Mahmarian JJ. Normal stress-only versus standard stress/rest myocardial perfusion imaging: Similar patient mortality with reduced radiation exposure. J Am Coll Cardiol 2010;55:221-30.CrossRef Chang SM, Nabi F, Xu J, Raza U, Mahmarian JJ. Normal stress-only versus standard stress/rest myocardial perfusion imaging: Similar patient mortality with reduced radiation exposure. J Am Coll Cardiol 2010;55:221-30.CrossRef
19.
Zurück zum Zitat Hussain N, Parker MW, Henzlova MJ, Duvall WL. Stress-first myocardial perfusion imaging. Cardiol Clin 2016;34:59-67.CrossRef Hussain N, Parker MW, Henzlova MJ, Duvall WL. Stress-first myocardial perfusion imaging. Cardiol Clin 2016;34:59-67.CrossRef
20.
Zurück zum Zitat Hudson HM, Larkin RS. Accelerated image reconstruction using ordered subsets projection data. IEEE Trans Med Imaging 1994;13:601-9.CrossRef Hudson HM, Larkin RS. Accelerated image reconstruction using ordered subsets projection data. IEEE Trans Med Imaging 1994;13:601-9.CrossRef
21.
Zurück zum Zitat Mathur S, Heller GV, Bateman TM, Ruffin R, Yekta A, Katten D, et al. Clinical value of stress-only Tc-99m SPECT imaging: Importance of attenuation correction. J Nucl Cardiol 2013;20:27-37.CrossRef Mathur S, Heller GV, Bateman TM, Ruffin R, Yekta A, Katten D, et al. Clinical value of stress-only Tc-99m SPECT imaging: Importance of attenuation correction. J Nucl Cardiol 2013;20:27-37.CrossRef
22.
Zurück zum Zitat Slomka PJ, Nishina H, Berman DS, Akincioglu C, Abidov A, Friedman JD, et al. Automated quantification of myocardial perfusion SPECT using simplified normal limits. J Nucl Cardiol 2005;12:66-77.CrossRef Slomka PJ, Nishina H, Berman DS, Akincioglu C, Abidov A, Friedman JD, et al. Automated quantification of myocardial perfusion SPECT using simplified normal limits. J Nucl Cardiol 2005;12:66-77.CrossRef
23.
Zurück zum Zitat Holly TA, Abbott BG, Al-Mallah M, Calnon DA, Cohen MC, DiFilippo FP, et al. Single photon-emission computed tomography. J Nucl Cardiol 2010;17:941-73.CrossRef Holly TA, Abbott BG, Al-Mallah M, Calnon DA, Cohen MC, DiFilippo FP, et al. Single photon-emission computed tomography. J Nucl Cardiol 2010;17:941-73.CrossRef
24.
Zurück zum Zitat Ichihara T, Ogawa K, Motomura N, Kubo A, Hashimoto S. Compton scatter compensation using the triple-energy window method for single-isotope and dual-isotope spect. J Nucl Med 1993;34:2216-21.PubMed Ichihara T, Ogawa K, Motomura N, Kubo A, Hashimoto S. Compton scatter compensation using the triple-energy window method for single-isotope and dual-isotope spect. J Nucl Med 1993;34:2216-21.PubMed
25.
Zurück zum Zitat Pretorius PH, King MA, Johnson KL, Yang Y, Wernick MN. Introducing polar map defined defects into normal cardiac perfusion SPECT slices using 3D respiratory and rigid-body motion projection. 13 th International Meeting of Fully Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine. New Port, RI; 2015. Pretorius PH, King MA, Johnson KL, Yang Y, Wernick MN. Introducing polar map defined defects into normal cardiac perfusion SPECT slices using 3D respiratory and rigid-body motion projection. 13 th International Meeting of Fully Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine. New Port, RI; 2015.
26.
Zurück zum Zitat Narayanan MV, King MA, Leppo J, Dahlberg S, Pretorius PH, Gifford HC. Optimization of regularization of attenuation and scatter-corrected Tc-99m cardiac SPECT studies for defect detection using hybrid images. IEEE Trans Nucl Sci 2001;48:785-9.CrossRef Narayanan MV, King MA, Leppo J, Dahlberg S, Pretorius PH, Gifford HC. Optimization of regularization of attenuation and scatter-corrected Tc-99m cardiac SPECT studies for defect detection using hybrid images. IEEE Trans Nucl Sci 2001;48:785-9.CrossRef
27.
Zurück zum Zitat Svane B. Polar presentation of coronary angiography and Thallium-201 single emission computed tomography. Sweden: Stockholm; 1990. Svane B. Polar presentation of coronary angiography and Thallium-201 single emission computed tomography. Sweden: Stockholm; 1990.
28.
Zurück zum Zitat Pretorius PH, Juan Ramon A, King MK, Konik A, Dahlberg ST, Parker M et al. Receiver operator characteristic confirmation of potential for radiation dose reduction with improved reconstruction for cardiac SPECT. IEEE Medical Imaging Conference. Atlanta, GA; 2017. Pretorius PH, Juan Ramon A, King MK, Konik A, Dahlberg ST, Parker M et al. Receiver operator characteristic confirmation of potential for radiation dose reduction with improved reconstruction for cardiac SPECT. IEEE Medical Imaging Conference. Atlanta, GA; 2017.
29.
Zurück zum Zitat Narayanan MV, Gifford HC, King MA, Pretorius PH, Farncombe TH, Bruyant P, et al. Optimization of iterative reconstruction of 99mTc Cardiac SPECT studies using numerical observers. IEEE Trans Nucl Sci 2002;49:2355-60.CrossRef Narayanan MV, Gifford HC, King MA, Pretorius PH, Farncombe TH, Bruyant P, et al. Optimization of iterative reconstruction of 99mTc Cardiac SPECT studies using numerical observers. IEEE Trans Nucl Sci 2002;49:2355-60.CrossRef
30.
Zurück zum Zitat Narayanan MV, Pretorius PH, Dahlberg ST, Leppo JA, Botkin N, Krasnow J, et al. Evaluation of scatter compensation strategies and their impact on human detection performance Tc-99m myocardial perfusion imaging. IEEE Trans Nucl Sci 2003;50:1522-7.CrossRef Narayanan MV, Pretorius PH, Dahlberg ST, Leppo JA, Botkin N, Krasnow J, et al. Evaluation of scatter compensation strategies and their impact on human detection performance Tc-99m myocardial perfusion imaging. IEEE Trans Nucl Sci 2003;50:1522-7.CrossRef
31.
Zurück zum Zitat Narayanan MV, King MA, Pretorius PH, Dahlberg ST, Spencer F, Simon E, et al. Human-observer receiver-operating-characteristic evaluation of attenuation, scatter, and resolution compensation strategies for Tc-99m myocardial perfusion imaging. J Nucl Med 2003;44:1725-34.PubMed Narayanan MV, King MA, Pretorius PH, Dahlberg ST, Spencer F, Simon E, et al. Human-observer receiver-operating-characteristic evaluation of attenuation, scatter, and resolution compensation strategies for Tc-99m myocardial perfusion imaging. J Nucl Med 2003;44:1725-34.PubMed
32.
Zurück zum Zitat Arsanjani R, Xu Y, Hayes SW, Fish M, Lemly M, Gerlach J, et al. Comparison of fully automated computer analysis and visual scoring for detection of coronary artery disease from myocardial SPECT in a large population. J Nucl Med 2013;54:221-8.CrossRef Arsanjani R, Xu Y, Hayes SW, Fish M, Lemly M, Gerlach J, et al. Comparison of fully automated computer analysis and visual scoring for detection of coronary artery disease from myocardial SPECT in a large population. J Nucl Med 2013;54:221-8.CrossRef
33.
Zurück zum Zitat Pretorius PH, King MA, Gifford HC, Dahlberg ST, Spencer F, Simon E, et al. Myocardial perfusion SPECT reconstruction: Receiver operating characteristic comparison of CAD detection accuracy of filtered backprojection reconstruction with all of the clinical imaging information available to readers and solely stress slices iteratively reconstructed with combined compensation. J Nucl Cardiol 2005;12:284-93.CrossRef Pretorius PH, King MA, Gifford HC, Dahlberg ST, Spencer F, Simon E, et al. Myocardial perfusion SPECT reconstruction: Receiver operating characteristic comparison of CAD detection accuracy of filtered backprojection reconstruction with all of the clinical imaging information available to readers and solely stress slices iteratively reconstructed with combined compensation. J Nucl Cardiol 2005;12:284-93.CrossRef
34.
Zurück zum Zitat Metz CE, Herman BA, Shen J-H. Maximum-likelihood estimation of ROC curves from continuously-distributed data. Stat Med 1998;17:1033-53.CrossRef Metz CE, Herman BA, Shen J-H. Maximum-likelihood estimation of ROC curves from continuously-distributed data. Stat Med 1998;17:1033-53.CrossRef
35.
Zurück zum Zitat Metz CE. Some practical issues of experimental design and data analysis in radiological ROC studies. Invest Radiol 1989;24:234-45.CrossRef Metz CE. Some practical issues of experimental design and data analysis in radiological ROC studies. Invest Radiol 1989;24:234-45.CrossRef
Metadaten
Titel
Retrospective fractional dose reduction in Tc-99m cardiac perfusion SPECT/CT patients: A human and model observer study
verfasst von
P. Hendrik Pretorius, PhD
Albert Juan Ramon, PhD
Michael A. King, PhD
Arda Konik, PhD
Seth T. Dahlberg, MD
Mathew W. Parker, MD
Naomi F. Botkin, MD
Karen L. Johnson, NMT
Yongyi Yang, PhD
Miles N. Wernick, PhD
Publikationsdatum
10.05.2019
Verlag
Springer International Publishing
Erschienen in
Journal of Nuclear Cardiology / Ausgabe 2/2021
Print ISSN: 1071-3581
Elektronische ISSN: 1532-6551
DOI
https://doi.org/10.1007/s12350-019-01743-7

Weitere Artikel der Ausgabe 2/2021

Journal of Nuclear Cardiology 2/2021 Zur Ausgabe

Ist Fasten vor Koronarinterventionen wirklich nötig?

Wenn Eingriffe wie eine Koronarangiografie oder eine Koronarangioplastie anstehen, wird häufig empfohlen, in den Stunden zuvor nüchtern zu bleiben. Ein französisches Forscherteam hat diese Maßnahme hinterfragt.

PET kann infarktgefährdete Koronararterien entdecken

04.06.2024 Koronare Herzerkrankung Nachrichten

Der Nachweis aktiver Plaques mittels 18F-Natriumfluorid-PET hilft nicht nur, infarktgefährdete Patienten, sondern auch infarktgefährdete Koronararterien zu erkennen. Von einer gezielten Behandlung vulnerabler Plaques ist man trotzdem weit entfernt.

GLP-1-Agonist Semaglutid wirkt kardio- und nephroprotektiv

03.06.2024 Semaglutid Nachrichten

Der GLP-1-Agonist Semaglutid hat in der FLOW-Studie bewiesen, dass sich damit die Progression chronischer Nierenerkrankungen bei Patienten mit Typ-2-Diabetes bremsen lässt. Auch in kardiovaskulärer Hinsicht war die Therapie erfolgreich.

Nach Herzinfarkt mit Typ-1-Diabetes schlechtere Karten als mit Typ 2?

29.05.2024 Herzinfarkt Nachrichten

Bei Menschen mit Typ-2-Diabetes sind die Chancen, einen Myokardinfarkt zu überleben, in den letzten 15 Jahren deutlich gestiegen – nicht jedoch bei Betroffenen mit Typ 1.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.