The relative trending accuracy of noninvasive continuous hemoglobin monitoring during hemodialysis in critically ill patients

verfasst von: Hiroshi Yamada, Minako Saeki, Junko Ito, Kazuhiro Kawada, Aya Higurashi, Hiromi Funakoshi, Kohji Takeda

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 1/2015

Einloggen, um Zugang zu erhalten

Abstract

The pulse CO-Oximeter (Radical-7; Masimo Corp., Irvine, CA) is a multi-wavelength spectrophotometric method for noninvasive continuous monitoring of hemoglobin (SpHb). Because evaluating the relative change in blood volume (ΔBV) is crucial to avoid hypovolemia and hypotension during hemodialysis, it would be of great clinical benefit if ΔBV could be estimated by measurement of SpHb during hemodialysis. The capability of the pulse CO-Oximeter to monitor ΔBV depends on the relative trending accuracy of SpHb. The purpose of the current study was to evaluate the relative trending accuracy of SpHb by the pulse CO-Oximeter using Crit-Line as a reference device. In 12 patients who received hemodialysis (total 22 sessions) in the intensive care unit, ΔBV was determined from SpHb. Relative changes in blood volume determined from SpHb were calculated according to the equation: ΔBV(SpHb) = [starting SpHb]/[current SpHb] − 1. The absolute values of SpHb and hematocrit measured by Crit-Line (CL-Hct) showed poor correlation. On the contrary, linear regression analysis showed good correlation between ΔBV(SpHb) and the relative change in blood volume measured by Crit-Line [ΔBV(CL-Hct)] (r = 0.83; P ≤ 0.001). Bland–Altman analysis also revealed good agreement between ΔBV(SpHb) and ΔBV(CL-Hct) (bias, −0.77 %; precision, 3.41 %). Polar plot analysis revealed good relative trending accuracy of SpHb with an angular bias of 4.1° and radial limits of agreement of 24.4° (upper) and −16.2° (lower). The results of the current study indicate that SpHb measurement with the pulse CO-Oximeter has good relative trending accuracy.
Literatur
1.
Zurück zum Zitat Bagshaw SM, Brophy PD, Cruz D, Ronco C. Fluid balance as a biomarker: impact of fluid overload on outcome in critically ill patients with acute kidney injury. Crit Care. 2008;12(4):169.PubMedCentralPubMedCrossRef Bagshaw SM, Brophy PD, Cruz D, Ronco C. Fluid balance as a biomarker: impact of fluid overload on outcome in critically ill patients with acute kidney injury. Crit Care. 2008;12(4):169.PubMedCentralPubMedCrossRef
2.
Zurück zum Zitat Payen D, de Pont AC, Sakr Y, Spies C, Reinhart K, Vincent JL. A positive fluid balance is associated with a worse outcome in patients with acute renal failure. Crit Care. 2008;12(3):R74.PubMedCentralPubMedCrossRef Payen D, de Pont AC, Sakr Y, Spies C, Reinhart K, Vincent JL. A positive fluid balance is associated with a worse outcome in patients with acute renal failure. Crit Care. 2008;12(3):R74.PubMedCentralPubMedCrossRef
3.
Zurück zum Zitat Locatelli F, Buoncristiani U, Canaud B, Kohler H, Petitclerc T, Zucchelli P. Haemodialysis with on-line monitoring equipment: tools or toys? Nephrol Dial Transpl. 2005;20(1):22–33.CrossRef Locatelli F, Buoncristiani U, Canaud B, Kohler H, Petitclerc T, Zucchelli P. Haemodialysis with on-line monitoring equipment: tools or toys? Nephrol Dial Transpl. 2005;20(1):22–33.CrossRef
4.
Zurück zum Zitat Garzoni D, Keusch G, Kleinoeder T, Martin H, Dhondt A, Cremaschi L, Tatsis E, Ibrahim N, Boer W, Kuehne S, Claus M, Zahn M, Schuemann E, Engelmann J, Hickstein H, Wojke R, Gauly A, Passlick-Deetjen J. Reduced complications during hemodialysis by automatic blood volume controlled ultrafiltration. Int J Artif Organs. 2007;30(1):16–24.PubMed Garzoni D, Keusch G, Kleinoeder T, Martin H, Dhondt A, Cremaschi L, Tatsis E, Ibrahim N, Boer W, Kuehne S, Claus M, Zahn M, Schuemann E, Engelmann J, Hickstein H, Wojke R, Gauly A, Passlick-Deetjen J. Reduced complications during hemodialysis by automatic blood volume controlled ultrafiltration. Int J Artif Organs. 2007;30(1):16–24.PubMed
5.
Zurück zum Zitat Steuer RR, Harris DH, Conis JM. A new optical technique for monitoring hematocrit and circulating blood volume: its application in renal dialysis. Dial Transplant. 1993;22(5):260–5. Steuer RR, Harris DH, Conis JM. A new optical technique for monitoring hematocrit and circulating blood volume: its application in renal dialysis. Dial Transplant. 1993;22(5):260–5.
6.
Zurück zum Zitat Shamir MY, Avramovich A, Smaka T. The current status of continuous noninvasive measurement of total, carboxy, and methemoglobin concentration. Anesth Analg. 2012;114(5):972–8.PubMedCrossRef Shamir MY, Avramovich A, Smaka T. The current status of continuous noninvasive measurement of total, carboxy, and methemoglobin concentration. Anesth Analg. 2012;114(5):972–8.PubMedCrossRef
7.
Zurück zum Zitat Critchley LA, Lee A, Ho AM. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010;111(5):1180–92.PubMedCrossRef Critchley LA, Lee A, Ho AM. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010;111(5):1180–92.PubMedCrossRef
8.
Zurück zum Zitat Dasselaar JJ, Huisman RM, DE Jong PE, Franssen CF. Relative blood volume measurements during hemodialysis: comparisons between three noninvasive devices. Hemodial Int. 2007;11(4):448–55. Dasselaar JJ, Huisman RM, DE Jong PE, Franssen CF. Relative blood volume measurements during hemodialysis: comparisons between three noninvasive devices. Hemodial Int. 2007;11(4):448–55.
9.
Zurück zum Zitat Lamhaut L, Apriotesei R, Combes X, Lejay M, Carli P, Vivien B. Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue(R) with automated laboratory hemoglobin measurement. Anesthesiology. 2011;115(3):548–54.PubMedCrossRef Lamhaut L, Apriotesei R, Combes X, Lejay M, Carli P, Vivien B. Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue(R) with automated laboratory hemoglobin measurement. Anesthesiology. 2011;115(3):548–54.PubMedCrossRef
10.
Zurück zum Zitat Macknet MR, Allard M, Applegate 2nd RL, Rook J. The accuracy of noninvasive and continuous total hemoglobin measurement by pulse CO-oximetry in human subjects undergoing hemodilution. Anesth Analg. 2010;111(6):1424–6. Macknet MR, Allard M, Applegate 2nd RL, Rook J. The accuracy of noninvasive and continuous total hemoglobin measurement by pulse CO-oximetry in human subjects undergoing hemodilution. Anesth Analg. 2010;111(6):1424–6.
11.
Zurück zum Zitat Colquhoun DA, Forkin KT, Durieux ME, Thiele RH. Ability of the Masimo pulse CO-Oximeter to detect changes in hemoglobin. J Clin Monit Comput. 2012;26(2):69–73.PubMedCrossRef Colquhoun DA, Forkin KT, Durieux ME, Thiele RH. Ability of the Masimo pulse CO-Oximeter to detect changes in hemoglobin. J Clin Monit Comput. 2012;26(2):69–73.PubMedCrossRef
12.
Zurück zum Zitat Frasca D, Dahyot-Fizelier C, Catherine K, Levrat Q, Debaene B, Mimoz O. Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients. Crit Care Med. 2011;39(10):2277–82.PubMedCrossRef Frasca D, Dahyot-Fizelier C, Catherine K, Levrat Q, Debaene B, Mimoz O. Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients. Crit Care Med. 2011;39(10):2277–82.PubMedCrossRef
13.
Zurück zum Zitat Critchley LA, Yang XX, Lee A. Assessment of trending ability of cardiac output monitors by polar plot methodology. J Cardiothorac Vasc Anesth. 2011;25(3):536–46.PubMedCrossRef Critchley LA, Yang XX, Lee A. Assessment of trending ability of cardiac output monitors by polar plot methodology. J Cardiothorac Vasc Anesth. 2011;25(3):536–46.PubMedCrossRef
14.
Zurück zum Zitat Giraud B, Frasca D, Debaene B, Mimoz O. Comparison of haemoglobin measurement methods in the operating theatre. Br J Anaesth. 2013;111(6):946–54.PubMedCrossRef Giraud B, Frasca D, Debaene B, Mimoz O. Comparison of haemoglobin measurement methods in the operating theatre. Br J Anaesth. 2013;111(6):946–54.PubMedCrossRef
15.
Zurück zum Zitat Mitra S, Chamney P, Greenwood R, Farrington K. The relationship between systemic and whole-body hematocrit is not constant during ultrafiltration on hemodialysis. J Am Soc Nephrol. 2004;15(2):463–9.PubMedCrossRef Mitra S, Chamney P, Greenwood R, Farrington K. The relationship between systemic and whole-body hematocrit is not constant during ultrafiltration on hemodialysis. J Am Soc Nephrol. 2004;15(2):463–9.PubMedCrossRef
16.
Zurück zum Zitat Booth J, Pinney J, Davenport A. Do changes in relative blood volume monitoring correlate to hemodialysis-associated hypotension? Nephron Clin Pract. 2011;117(3):c179–83.PubMedCrossRef Booth J, Pinney J, Davenport A. Do changes in relative blood volume monitoring correlate to hemodialysis-associated hypotension? Nephron Clin Pract. 2011;117(3):c179–83.PubMedCrossRef
17.
Zurück zum Zitat Miller RD, Ward TA, Shiboski SC, Cohen NH. A comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery. Anesth Analg. 2011;112(4):858–63.PubMedCrossRef Miller RD, Ward TA, Shiboski SC, Cohen NH. A comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery. Anesth Analg. 2011;112(4):858–63.PubMedCrossRef
Metadaten
Titel
The relative trending accuracy of noninvasive continuous hemoglobin monitoring during hemodialysis in critically ill patients
verfasst von
Hiroshi Yamada
Minako Saeki
Junko Ito
Kazuhiro Kawada
Aya Higurashi
Hiromi Funakoshi
Kohji Takeda
Publikationsdatum
01.02.2015
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 1/2015
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-014-9574-6

Weitere Artikel der Ausgabe 1/2015

Journal of Clinical Monitoring and Computing 1/2015 Zur Ausgabe

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.