Skip to main content
Erschienen in: Journal of Clinical Monitoring and Computing 1/2015

Open Access 01.02.2015 | Original Research

Thoracic impedance measures tissue characteristics in the vicinity of the electrodes, not intervening lung water: implications for heart failure monitoring

verfasst von: Christopher J. Charles, Miriam T. Rademaker, Iain C. Melton, Dan Gutfinger, Neal L. Eigler, Fujian Qu, Richard W. Troughton

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 1/2015

Abstract

The rationale for intrathoracic impedance (Z) detection of worsening heart failure (HF) presupposes that changes in Z reflect changes in pulmonary congestion, but is confounded by poor specificity in clinical trials. We therefore tested the hypothesis that Z is primarily affected by tissue/water content in proximity to electrodes rather than by lung water distribution between electrodes through the use of a new computational model for deriving the near-field impedance contributions from the various electrodes. Six sheep were implanted with a left atrial pressure (LAP) monitor and a cardiac resynchronization therapy device which measured Z from six vectors comprising of five electrodes. The vector-based Z was modelled as the summation of the near-field impedances of the two electrodes forming the vector. During volume expansion an acute increase in LAP resulted in simultaneous reductions in the near-field impedances of the intra-cardiac electrodes, while the subcutaneous electrode showed several hours of lag (all p < 0.001). In contrast, during the simulated formation of device-pocket edema (induced by fluid injection) the near-field impedance of the subcutaneous electrode had an instantaneous response, while the intra-cardiac electrodes had a minimal inconsistent response. This study suggests that the primary contribution to the vector based Z is from the tissue/water in proximity to the individual electrodes. This novel finding may help explain the limited utility of Z for detecting worsening HF.

1 Introduction

Despite treatment advances and more intensive clinical monitoring, heart failure (HF) remains one of the most common causes for hospitalization and is associated with high morbidity, mortality and economic costs [1]. The increasing use of implantable cardioverter defibrillators (ICD) and cardiac resynchronization therapy (CRT) devices in patients with HF makes it possible to remotely monitor intrathoracic impedance (Z) in combination with other physiological parameters [2]. Z measurements using the right ventricular (RV) shocking coil to the device case vector (RVcoil-Case) are traditionally thought to represent the resistance to electrical flow across a field extending through the lungs, such that changes in Z can be interpreted to reflect changes in thoracic fluid volume and pulmonary congestion [3, 4]. Based on this premise several studies have shown that it is possible to use Z to predict and potentially reduce hospitalizations for acute decompensated HF (ADHF) [5, 6]. However, because of the limited specificity and sensitivity of Z monitoring when using the RVcoil-Case vector [5, 7], more recent studies have explored the use of alternative and combination vectors from CRT leads [810].
Pre-clinical studies using multiple vectors have shown that each vector changes at a different rate and magnitude in response to the development of pacing-induced HF [9, 10]. The left ventricular (LV) ring electrode to device case vector (LVring-Case) demonstrated the fastest rate and greatest magnitude of change, and correlated well with left atrial pressure (LAP). A recent clinical study reported that a combination of Z vectors, rather than Z from a single vector can successfully identify increases in thoracic fluid volume that will culminate in ADHF. Using a combination vector from either CRT or ICD leads showed improved sensitivity and a lower false positive rate compared with the single RVcoil-Case vector [11].
In order to more effectively exploit the availability of Z from multiple vectors, we sought to gain better insight into what is being measured by these vectors. The electric current flowing between two electrodes forming a vector does not necessarily travel along a straight line across the lungs, but more likely along pathways of lesser resistance. We, therefore, hypothesized that the most significant contribution to the measured Z is from the electrode-tissue interface where there is high impedance to electrical flow, and that beyond this interface the contribution to the measured Z is negligible [12]. This is because beyond the electrode-tissue interface the current finds its way to flow through low resistance structures, such as blood vessels, that do not have a significant contribution to the measured Z. If Z primarily depends on the electrode-tissue interface, then it provides a measure of the local tissue characteristics surrounding the two electrodes rather than the characteristics of tissues located farther away between the electrodes. This new interpretation may help explain why recent clinical trials have shown that monitoring of Z has limited effectiveness in detecting ADHF [13, 14], particularly in the early months post-device implant when there are significant non-HF related changes occurring at the electrode-tissue interface. To test the hypothesis that Z primarily reflects local tissue characteristics we evaluated a new computational model for deriving the near-field impedance contributions from the various electrodes and conducted a pre-clinical experiment to illustrate how near-field impedances may be monitored at multiple sites.

2 Methods

The study protocol was approved by the Animal Ethics Committee of the University of Otago, Christchurch, New Zealand.

2.1 Implantation Procedure and Devices

Under general anesthesia (induced by 15 mg/kg thiopentone; maintained by halothane/nitrous oxide inhalation) six sheep were implanted via the right jugular vein with an LAP monitor (HeartPOD™ ISL) and three standard transvenous bipolar pacing/defibrillation leads with the tip electrodes fixed in the right atrial (RA) appendage, RV apex and LV epicardial region via the coronary sinus. The leads were attached to a CRT device (Promote™ RF) that was implanted subcutaneously over the left chest wall. The CRT device provided Z measurements along six vectors: V1 = LVring-Case; V2 = RVring-Case; V3 = RAring-Case; V4 = RVcoil-Case; V5 = LVring-RVring; and V6 = LVring-RAring [9, 10]. The LAP monitor was implanted into the left atrium via a trans-septal catheterization, and provided direct LAP waveform measurements via telemetry communication [15, 16]. All implanted devices were manufactured by St Jude Medical, Implantable Electronic Systems Division in Sylmar, CA, USA. Post-operatively, animals recovered for a minimum of 5 weeks to allow the Z measurements to stabilize.

2.2 Acute study protocols

Following the stabilization period, sheep were studied on separate days as follows:

2.2.1 Intravascular volume expansion in normal sheep (pre-pacing)

Sheep received a 4-h intravenous (IV) infusion of Dextran (initial dose 15 ml/kg over 30 min, then titrated to maintain LAP at approximately 15 mmHg above baseline values).

2.2.2 Induction of Congestive HF

Sheep underwent 7 days of rapid RV pacing at 190 bpm to induce a state of moderate stable congestive HF. This rate of RV pacing consistently raised LAP levels in this animal model by approximately 10 mmHg [17].

2.2.3 Intravascular volume expansion during congestive HF

Sheep received a 4-h IV infusion of Dextran (as above) in the HF state. Dextran was titrated to raise LAP by approximately 15 mmHg (from HF baseline levels of 16–17 mmHg to approximately 31–32 mmHg).

2.2.4 Device pocket edema

To simulate the formation of edema within the device pocket 250 mL of 0.9 % saline was rapidly injected via an angiocath catheter into the subcutaneous pocket containing the CRT device case. An additional 250 mL saline bolus was injected in 5/6 sheep 2 h later.

2.3 Data Acquisition and Processing

LAP measurements were recorded every 5 min throughout the acute study protocols as previously described [15, 16]. Z measurements were automatically collected through the CRT device along all six vectors (V1 through V6) at 7.5 min intervals throughout the acute study protocols. Using equivalent circuit equations [18], the measured Z along each vector was modelled as the sum of the near-field contributions from the two measuring electrodes to produce a set of six linear equations (Fig. 1). The set of equations was solved algebraically to determine the near-field impedance contribution (denoted by ZE or the electrode name) associated with each of the five electrodes. As outlined in Fig. 1, the ZE associated with the LVring electrode was computed using the equations corresponding to two different impedance triangles to yield LV1 and LV2 which were then averaged. The difference between LV1 and LV2 determined the measurement error associated with the computational model.

2.4 Statistics

All results are expressed as mean ± SEM unless otherwise specified. For each acute study the effect on LAP, Z and ZE parameters were assessed using one-way repeated measures analysis of variance (ANOVA). A value of p < 0.05 was considered statistically significant.

3 Results

3.1 Near-field impedance measurement validation

Near-field impedances (ZE) and the relative contributions from each of the five electrodes toward the vector-based Z measurements are summarized in Table 1. As expected, the highest ZE was associated with the smaller sized electrodes (LVring, RVring and RAring), while the lowest ZE was associated with the larger sized electrodes (Case and RVcoil). Electrode size also influenced the relative contributions from each of the electrodes towards the measured Z. For a vector between any ring electrode (LVring, RVring, and RAring) and the Case electrode (V1 through V3), most of the contribution to Z was found to be from the ring electrode (77-87 %), while only 13–23 % of the contribution was from the Case. In contrast, for the RVcoil to Case vector (V4) most of the contribution to Z was found to be from the subcutaneous Case electrode (69 %), with only 31 % from the intra-cardiac RVcoil electrode. For the intra-cardiac vectors that are between two ring electrodes (V5 and V6) the contributions from the two electrodes was found to be similar. Thus, it appears that for a vector with different electrode sizes, the smaller electrode or the electrode that is surrounded by more scar tissue will be the one that is most influencing any changes in the vector-based Z measurement.
Table 1
Mean impedance measurements with relative contribution of near-field ZE to vector-based Z. Data averaged across all sheep (N = 6) for all measurements over entire duration of implants [100 ± 22 (SD) days]
Vector
Mean ± SD Z (Ω)
Contribution from 1st electrode
Contribution from 2nd electrode
Electrode
Mean ± SD ZE (Ω)
V1: LVring-Case
306 ± 53
87 % LVring
13 % Case
LVring
266 ± 50
V2: RVring-Case
177 ± 29
77 % RVring
23 % Case
RVring
137 ± 26
V3: RAring-Case
254 ± 28
84 % RAring
16 % Case
RAring
214 ± 26
V4: RVcoil-Case
58 ± 6
31 % RVcoil
69 % Case
Case
40 ± 5
V5: LVring-RVring
403 ± 50
66 % LVring
34 % RVring
RVcoil
18 ± 5
V6: LVring-RAring
480 ± 58
55 % LVring
45 % RAring
  
    
LV1
268 ± 50
    
LV2
265 ± 50
The error in computing the near-field impedance for the LVring electrode (Fig. 1) was found to be extremely small at 2.9 ± 3.8 Ω (~1 %). This small error in combination with the observations regarding electrode size influencing ZE both validate the computational model used for deriving the near-field impedance contributions to the vector-based Z measurements.

3.2 Intravascular volume expansion in normal sheep (pre-pacing)

The total volume of Dextran administered over the 4 h infusion period ranged between 3,500 and 4,500 mL. LAP increased from 7 ± 1 to 21 ± 1.4 mmHg by 30 min and remained stable while Dextran was being infused (Fig. 2). Once the infusion terminated, LAP gradually fell over 60 min, but remained 2–3 mmHg above baseline levels for the remainder of the recording period. In comparison, after a brief delay (30 min), LVring-Case Z fell (p < 0.001), reaching a nadir at 285 min and remaining low. ZE from LVring (p < 0.001) showed a similar response, achieving a nadir after 270 min. RVring-Case Z fell without delay reaching a nadir at 265 min (p < 0.001). In contrast, RVcoil-Case Z demonstrated a delayed fall (90 min), and did not reach nadir till 375 min (p < 0.001). The RVring, RVcoil and RAring ZE all fell immediately and abruptly (all p < 0.001) mirroring the acute change in LAP, and reaching nadirs at 45, 30 and 30 min, respectively. Despite the observation that LAP fell within 60 min following Dextran infusion, the intra-cardiac ZE measurements remained suppressed for the duration of the recording period. Case ZE changes were less consistent, with the level actually rising initially and ultimately reducing with significant lag.

3.3 Induction of congestive HF

Pacing at 190 bpm induced a moderate state of HF as indicated by rapid and significant increases in LAP, rising from a pre-pacing baseline of 7 ± 1–16 ± 1 mmHg within a few hours of pacing, before plateauing (p < 0.001). Pacing induced the expected decreases in vector-based Z, falling from pre-pacing levels of 310 ± 21–257 ± 21 Ω, 186 ± 12–162 ± 10 Ω and 61 ± 1–54 ± 2 Ω for LVring-Case, RVring-Case and RVcoil-Case, respectively (all p < 0.001). Similarly, pacing produced falls in the ZE for LVring (269 ± 19–220 ± 19 Ω, p < 0.001), RVring (145 ± 12–125 ± 19 Ω, p = 0.004) and Case (41 ± 2.1 vs. 37 ± 1.9 Ω, p = 0.014) electrodes. The ZE of the RVcoil (20.2 ± 2.1–17.5 ± 2.1 Ω, p = 0.07) and RAring (215 ± 11–203 ± 11 Ω, p = 0.16) electrodes also tended to decline.

3.4 Intravascular volume expansion during congestive HF

The total volume of Dextran administered over the infusion period was 2,900–4,000 mL. Acute volume loading increased LAP from 16 ± 1 to 31 ± 1.2 mmHg by 30 min, with pressures remaining stable for the duration of the infusion (Fig. 3). LAP gradually reduced following termination of Dextran infusion, but remained above pre-Dextran infusion levels. Both LVring-Case and RVcoil-Case vectors showed a small but consistent rise (both p < 0.001) in response to Dextran (Fig. 3), with similar rises observed for ZE of the LVring (p < 0.001) and Case (p < 0.001). In contrast, RVring-Case vector-based Z and ZE from RVring, RAring and RVcoil all showed prompt and consistent falls (all p < 0.001), reaching nadirs by 37.5 min and remaining depressed for the duration of the recording period.

3.5 Device pocket edema

There was a small but consistent elevation in LAP (~2 mmHg; p < 0.001) during the study day which appeared unrelated to fluid injection (Fig. 4). All three vector-based Z showed an immediate decline (all p < 0.001) in response to fluid injection into the device pocket, and remained lower than baseline. ZE from the Case showed a prompt and dramatic fall in response to pocket fluid injection (p < 0.001), dropping from ~43 to ~30 Ω following the first injection, and then to ~27 Ω following the second injection. In contrast, ZE associated with the other four electrodes showed minimal and inconsistent changes unrelated to fluid injection.

4 Discussion

In this study we demonstrate for the first time that changes in vector-based Z measurements may readily develop in response to changes occurring at the electrode-tissue interface, thereby providing a possible mechanism that explains why Z monitoring has limited sensitivity and specificity in detecting ADHF [13]. The hypothesis that Z primarily reflects local tissue characteristics suggests that non-HF events, such as healing post-implant, lead migration/dislodgement, and device-pocket infection, can easily confound the interpretation of changes in Z measurements [4, 14]. Thus, in the current study we evaluated changes in both vector-based Z and ZE responses during rapid changes in LAP induced by volume and non-volume-dependent mechanisms. Volume loading with dextran is a standard physiological model designed to acutely increase intravascular and intracardiac volumes [19, 20]. The rate and duration of dextran infusion in the current study was sufficient to increase intravascular and intracardiac volumes with an approximate 15 mmHg rise in LAP, but is unlikely to have caused pulmonary edema. Rapid ventricular pacing is a well established animal model of congestive HF which consistently induces significant sodium and water retention, changes in intracardiac volumes and pressures and pulmonary edema [17]. This model has been previously used in our lab to study vector-based impedance changes with the onset and offset of congestive HF, thereby modelling ADHF decompensation and recovery [10]. Results from the current study confirm that pacing induced HF not only reduces all vector-based impedances, as previously shown [10], but also consistently reduces ZE derived from each of the electrodes.
Results from the current study support our hypothesis that Z primarily reflects local tissue characteristics by demonstrating that intrathoracic impedance measurements reflect the local tissue and fluid conditions surrounding each of the electrodes, rather than the characteristics of tissues located farther away in the region between the measuring electrodes. The highest near-field impedance (ZE) was associated with the smaller-sized LVring, RVring and RAring electrodes which had greater resistance to electrical flow at the electrode-tissue interface (Table 1). The lowest ZE was associated with the Case and RVcoil electrodes, consistent with these electrodes being larger in size with less resistance to electrical flow. The LVring electrode had the highest ZE, consistent with this electrode being surrounded by more tissue and less blood.
To construct the mathematical model we hypothesized that the current flows along a pathway from the first electrode to the second electrode which includes three components; (1) the tissues adjacent to the first electrode; (2) the tissues and/or blood located farther away between the two electrodes; and (3) the tissues adjacent to the second electrode. The measured impedance between the first electrode and the second electrode therefore can be modelled as the summation of the impedance associated with each of these pathway components. We also hypothesized that the contribution to the measured impedance between the first and the second electrode is greatest from the 1st and 3rd components of the pathway, and that the contribution from the 2nd component is significantly lower such that it may be ignored. We believe this hypothesis to be correct because the electrical current will find its way to flow through a pathway between the electrodes that has the lowest impedance to electrical flow, such as along blood vessels. Thus, the mathematical model for the measured impedance between two electrodes was simplified to consist of only the 1st and 3rd components. Since we had six measurements of vector based impedance from five electrodes we were able to compute the contribution from the LVring electrode in more than one way (LV1 and LV2) which allowed us to verify our hypothesis that the impedance associated with the 2nd component pathway can be ignored.
The most striking example of impedance reflecting the local tissue and fluid conditions surrounding an electrode was seen in the device pocket edema study where, although vector-based Z (V1, V2, and V4) fell in response to the injection of fluid into the device pocket, near-field computation demonstrated that the marked and precipitous fall in the Case ZE contributed most, if not all, to the decrease in the vector-based Z. This can be seen in Fig. 5a where the falls in Z for V1, V2, and V4 are plotted at 30 min following saline injection into the pocket. The reduction in Z for V1 and V4 is contributed almost exclusively by the fall in ZE of the Case electrode, with little if any contribution from the ZE of the LVring or RVcoil electrodes, respectively. The large accumulation of fluid surrounding the Case resulted in little discernable change in LAP or ZE for any of the other implanted electrodes located far away from the device pocket. A similar response pattern in the measured Z is classically seen whenever a pulse generator is replaced, which results in the formation of local edema within the device pocket during the first 24 h, followed by the normal phases of tissue healing occurring over the subsequent weeks to months. Thus, device pocket edema induces significant reductions in the measured Z that may be falsely interpreted [4]. However, with the ability to remove the contribution of the Case electrode from the measured vector-based Z, it is possible to monitor impedance without it being subjected to otherwise confounding changes that are occurring within the device pocket.
The hypothesis that Z primarily reflects local tissue characteristics is further supported by the intravascular volume expansion studies performed both in normal (pre-pacing) and HF sheep which induced a significant increase in LAP that was promptly tracked by corresponding reductions in the ZE measurements from the intra-cardiac (RVring, RAring and RVcoil) electrodes. The acute and the delayed responses to intravascular volume expansion in the normal (pre-pacing) sheep are illustrated in Fig. 5b and c, respectively. At 30 min following commencement of volume expansion, which corresponds to the time when LAP first peaked, there is a dramatic fall in Z for V2 and V4, whereas for V1 there is a slight unexpected increase (Fig. 5b). Near-field computation demonstrates that the acute reduction in Z for V2 and V4 is contributed entirely by a fall in the ZE for the RVring and RVcoil electrodes, respectively. Because the RVring and RVcoil electrodes are located within the RV chamber, a rapid increase in chamber volume causes more fluid to surround the electrodes and less tissue to be in contact with the electrodes, such that the near-field contribution to the measured vector-based Z is proportionally reduced.
Following the 30 min time point the Z for V1 has a steady decrease (Fig. 2), which is occurring over the time period when the infused fluid is being redistributed to other compartments. At the 300 min time point (Fig. 5c) near-field computation demonstrates that virtually all of the reduction in Z for V1 is contributed by the fall in the ZE for the LVring electrode. Although the LVring electrode is implanted within an epicardial blood vessel, its delayed response suggests that the LVring electrode is not in direct communication with the RV blood pool. This may be secondary to possible formation of scar tissue within the distal coronary venous implant site (confirmed by autopsy in some animals) causing obstruction to blood flow. Importantly, ZE from the LVring electrode was found to be sensitive to congestion within the adjacent epicardial fat, as demonstrated post volume expansion in canine studies in our lab (unpublished observations), and to pericardial fluid volume and the degree of contact between the LVring electrode and the pericardial sac [21]. Thus, an increase in the degree of tissue contact between the LVring electrode and the pericardial sac as a consequence of an expanding cardiac chamber volume may explain the unexpected initial increase observed in response to acute volume infusion, while the later development of epicardial fat edema may explain the delayed response. A change in the degree of contact between the LVring electrode and the pericardial sac may also occur in response to changes in posture or pericardial fluid volume irrespective of pulmonary congestion, and may result in false interpretations of the measured impedance as previously reported [21]. However, the influence of free flowing pericardial fluid volume and posture may be less critical in patients with a scarred pericardial space following open heart surgery, or an episode of pericarditis.
There were variations in the recovery pattern of the measured ZE following cessation of volume loading that depended on the electrode implant location. All five electrodes had a much slower recovery rate in ZE (>12 h) compared to LAP (~3 h). The intra-cardiac right-sided electrodes demonstrated a faster recovery compared to the LVring and Case electrodes. The delay in recovery of impedance in comparison to LAP is consistent with previous reported findings [10], and is likely a consequence of the longer time required to restore the local fluid volume conditions surrounding the various electrodes, particularly when located outside the intravascular space. Similarly, termination of rapid ventricular pacing caused a precipitous acute fall in LAP, followed by a gradual recovery over the subsequent 12 h. This was in contrast to the significantly slower recovery seen in ZE (~3 days—data not shown), which is consistent with previously reported data for ZV [10]. The lag seen between ZE and LAP is likely a consequence of ZE being more a reflection of the local volume and corresponding tissue contact rather than pressure.

4.1 Illustrative case studies

1.
Micro-dislodgement. Figure 6 provides an example of the Z measurements acquired from a canine surgically prepared in a similar manner to the sheep in this study [9]. This animal developed a micro-dislodgement of the LV lead during the third week post-implant. The trend data demonstrates that during the first 24 h post-implant there is a significant decrease in the measured Z across all six vectors, which is consistent with the development of edema in the vicinity of all five electrodes. During the subsequent weeks the Z measurements gradually increased as the local edema resolved. However, there is a sudden disturbance in the measured Z on day 18 post-implant seen across the three vectors having the LVring in common (V1, V5, and V6). Examination of the ZE shows that only the LVring electrode recorded a significant increase, with no change noted in the contribution from the other four electrodes. The ability to isolate the micro-dislodgement event into a single electrode additionally supports the computational model used to derive the near-field impedance contributions.
 
2.
Hospitalization for ADHF. An example of ZE data acquired from a 66 year-old patient with ischemic cardiomyopathy is shown as a trend plot over time in Fig. 7. This patient provided informed consent to participate in the clinical trial reported by Binkley and colleagues [11]. Approximately 2 months following implant, the patient developed an episode of persistent atrial fibrillation that resulted in hospitalization for ADHF. During hospitalization, the patient was treated with intravenous diuretic therapy, and underwent an AV-node ablation procedure (Fig. 7). In response to this treatment, the ZE from the RVring and Case electrodes increased dramatically, while the rise for the LVring ZE was more gradual and continued post-hospital discharge. Interestingly, the ZE from the RVcoil and RAring electrodes had more noise and measurement artifact, and did not coincide with the clinical course. This case illustrates that the response seen in ZE during the development and recovery from ADHF varies with the electrode location and the local tissue characteristics.
 

5 Summary

Taken together, the results of the present study validate the computational model used to derive the near-field impedances (ZE), and demonstrate that vector-based Z primarily depends on the impedance from the tissues in the vicinity of the measuring electrodes, and therefore, reflects local tissue characteristics. This finding may explain why vector-based Z can display changes that do not coincide with pulmonary congestion, particularly in the early months post-implant when there are significant changes occurring at the electrode-tissue interface. This is consistent with the results from a recent clinical trial which demonstrate that the positive predictive value of a Z monitoring algorithm is particularly low in the initial 6 months post-implant, but subsequently improves over time [13]. Instabilities post-implant are more likely to occur with Z that are associated with a passive fixation lead, such as an LV pacing lead, where there is a greater chance for a micro-dislodgement to occur (Fig. 6). Therefore, it may be preferable to allow more time for the electrode-tissue interface to stabilize before relying on Z measurements for clinical decision making, or to rely on Z vectors that are derived from an active fixation lead (RVcoil-Case and RVring-Case). Once the electrode-tissue interface is stabilized, this study demonstrates that the ZE associated with the intra-cardiac electrodes (RVring, RAring, and RVcoil) respond faster than the ZE associated with the LVring and Case electrode to an expanding intra-cardiac chamber volume. The ZE associated with the LVring and Case electrodes are additionally influenced by changes occurring within the pericardial space or device pocket, which may not coincide with pulmonary congestion. Therefore, relying on the ZE associated with the intra-cardiac electrodes for clinical decision-making may be more advantageous than other electrodes or Z vectors.

6 Conclusion

We have reported for the first time a validated method for determining the relative contributions to the measured vector-based impedance from the various electrodes, and have demonstrated that vector-based impedance measurements predominantly reflect physical phenomena occurring in the vicinity of the measuring electrodes rather than farther away within the pulmonary parenchyma. Thus, the near-field computational model offers a new paradigm for examining impedance data from multi-lead devices which may prove clinically-useful in monitoring changes at specific anatomical locations in close proximity to the individual recording electrodes. The results of this study help explain why vector-based impedance measurements may display changes that do not necessarily correlate with the presence of pulmonary congestion, and support the hypothesis that intrathoracic impedance actually measures local tissue characteristics rather than intervening lung water.

Acknowledgments

We are grateful to the staff of the University of Otago, Christchurch Animal Research Area for animal care. Supported in part by a research grant from St Jude Medical, and by The Health Research Council of New Zealand.

Conflict of interest

Eigler is an inventor of the HeartPOD™ device; Gutfinger, Eigler and Qu are all employees of St Jude Medical which has commercial interests in the HeartPOD™ and Promote™ devices.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Literatur
1.
Zurück zum Zitat American Heart Association Statistics Committee and Stroke Statistics Subcommittee. Heart disease and stroke statistics-2008 update: a report from the American heart association statistics committee and stroke statistics subcommittee. Circulation. 2008;117:e25–146.CrossRef American Heart Association Statistics Committee and Stroke Statistics Subcommittee. Heart disease and stroke statistics-2008 update: a report from the American heart association statistics committee and stroke statistics subcommittee. Circulation. 2008;117:e25–146.CrossRef
2.
Zurück zum Zitat Merchant FM, Dec GW, Singh JM. Implantable sensors for heart failure. Circ Arrhythm Electrophysiol. 2010;3:657–67.PubMedCrossRef Merchant FM, Dec GW, Singh JM. Implantable sensors for heart failure. Circ Arrhythm Electrophysiol. 2010;3:657–67.PubMedCrossRef
3.
Zurück zum Zitat Yu CM, Wang L, Chau E, et al. Intrathoracic impedance monitoring in patients with heart failure: correlation with fluid status and feasibility of early warning preceding hospitalization. Circulation. 2005;112:841–8.PubMedCrossRef Yu CM, Wang L, Chau E, et al. Intrathoracic impedance monitoring in patients with heart failure: correlation with fluid status and feasibility of early warning preceding hospitalization. Circulation. 2005;112:841–8.PubMedCrossRef
4.
Zurück zum Zitat Wang L. Fundamentals of intrathoracic impedance monitoring in heart failure. Am J Cardiol. 2007;99(suppl):3G–10G.PubMedCrossRef Wang L. Fundamentals of intrathoracic impedance monitoring in heart failure. Am J Cardiol. 2007;99(suppl):3G–10G.PubMedCrossRef
5.
Zurück zum Zitat Ypenburg C, Bax JJ, van der Wall EE, Schalij MJ, van Erven L. Intrathoracic Impedance monitoring to predict decompensated heart failure. Am J Cardiol. 2007;99:554–7.PubMedCrossRef Ypenburg C, Bax JJ, van der Wall EE, Schalij MJ, van Erven L. Intrathoracic Impedance monitoring to predict decompensated heart failure. Am J Cardiol. 2007;99:554–7.PubMedCrossRef
6.
Zurück zum Zitat Catanzarita D, Lunati M, Landolina M, et al. Monitoring intrathoracic impedance with an implantable defibrillator reduces hospitalizations in patients with heart failure. PACE. 2009;32:363–70.CrossRef Catanzarita D, Lunati M, Landolina M, et al. Monitoring intrathoracic impedance with an implantable defibrillator reduces hospitalizations in patients with heart failure. PACE. 2009;32:363–70.CrossRef
7.
Zurück zum Zitat Vollmann D, Nagele H, Schauerte P, et al. Clinical utility of intrathoracic impedance monitoring to alert patients with an implanted device of deteriorating chronic heart failure. Eur Heart J. 2007;28:1835–40.PubMedCrossRef Vollmann D, Nagele H, Schauerte P, et al. Clinical utility of intrathoracic impedance monitoring to alert patients with an implanted device of deteriorating chronic heart failure. Eur Heart J. 2007;28:1835–40.PubMedCrossRef
8.
Zurück zum Zitat Maines M, Landolina M, Lunati M, et al. Intrathoracic and ventricular impedances are associated with changes in ventricular volume in patients receiving defibrillators for CRT. PACE. 2010;33:64–73.PubMedCrossRef Maines M, Landolina M, Lunati M, et al. Intrathoracic and ventricular impedances are associated with changes in ventricular volume in patients receiving defibrillators for CRT. PACE. 2010;33:64–73.PubMedCrossRef
9.
Zurück zum Zitat Khoury DS, Naware M, Siou J, et al. Ambulatory monitoring of congestive heart failure by multiple bioelectric impedance vectors. J Am Coll Cardiol. 2009;53:1075–81.PubMedCentralPubMedCrossRef Khoury DS, Naware M, Siou J, et al. Ambulatory monitoring of congestive heart failure by multiple bioelectric impedance vectors. J Am Coll Cardiol. 2009;53:1075–81.PubMedCentralPubMedCrossRef
10.
Zurück zum Zitat Rademaker MT, Charles CJ, Melton IC, et al. Monitoring heart failure: comparison of left atrial pressure with intrathoracic impedance and natriuretic peptide measurements in an experimental model of ovine heart failure. Clin Sci. 2011;120:207–17.PubMedCentralPubMedCrossRef Rademaker MT, Charles CJ, Melton IC, et al. Monitoring heart failure: comparison of left atrial pressure with intrathoracic impedance and natriuretic peptide measurements in an experimental model of ovine heart failure. Clin Sci. 2011;120:207–17.PubMedCentralPubMedCrossRef
11.
Zurück zum Zitat Binkley PF, Porterfied JG, Porterfield LM, et al. Feasibility of using multivector impedance to monitor pulmonary congestion in heart failure patients. J Interv Card Electrophysiol. 2012;35:197–206.PubMedCrossRef Binkley PF, Porterfied JG, Porterfield LM, et al. Feasibility of using multivector impedance to monitor pulmonary congestion in heart failure patients. J Interv Card Electrophysiol. 2012;35:197–206.PubMedCrossRef
12.
Zurück zum Zitat Wittkampf FHM, Nakagawa H. RF catheter ablation: lessons on lesions. PACE. 2006;29:1285–97.PubMedCrossRef Wittkampf FHM, Nakagawa H. RF catheter ablation: lessons on lesions. PACE. 2006;29:1285–97.PubMedCrossRef
13.
Zurück zum Zitat Conraads VM, Tavazzi L, Santini M, et al. Sensitivity and positive predictive value of implantable intrathoracic impedance monitoring as a predictor of heart failure hospitalizations: the SENSE-HF trial. Eur Heart J. 2011;32:2266–73.PubMedCrossRef Conraads VM, Tavazzi L, Santini M, et al. Sensitivity and positive predictive value of implantable intrathoracic impedance monitoring as a predictor of heart failure hospitalizations: the SENSE-HF trial. Eur Heart J. 2011;32:2266–73.PubMedCrossRef
14.
Zurück zum Zitat Van Veldhuisen DJ, Braunschweig F, Conraads V, et al. Intrathoracic impedance monitoring, audible patient alerts, and outcome in patients with heart failure. Circulation. 2011;124:1719–26.PubMedCrossRef Van Veldhuisen DJ, Braunschweig F, Conraads V, et al. Intrathoracic impedance monitoring, audible patient alerts, and outcome in patients with heart failure. Circulation. 2011;124:1719–26.PubMedCrossRef
15.
Zurück zum Zitat Ritzema J, Melton IC, Richards AM, et al. Direct left atrial pressure monitoring in ambulatory heart failure patients: initial experience with a new permanent implantable device. Circulation. 2007;116:2952–9.PubMedCrossRef Ritzema J, Melton IC, Richards AM, et al. Direct left atrial pressure monitoring in ambulatory heart failure patients: initial experience with a new permanent implantable device. Circulation. 2007;116:2952–9.PubMedCrossRef
16.
Zurück zum Zitat Ritzema J, Troughton R, Melton I, et al. Physician-directed patient self-management of left atrial pressure in advance chronic heart failure. Circulation. 2010;121:1086–95.PubMedCrossRef Ritzema J, Troughton R, Melton I, et al. Physician-directed patient self-management of left atrial pressure in advance chronic heart failure. Circulation. 2010;121:1086–95.PubMedCrossRef
17.
Zurück zum Zitat Rademaker MT, Charles CJ, Espiner EA, Frampton CM, Nicholls MG, Richards AM. Natriuretic peptide responses to acute and chronic ventricular pacing in sheep. Am J Physiol. 1996;270:H594–602.PubMed Rademaker MT, Charles CJ, Espiner EA, Frampton CM, Nicholls MG, Richards AM. Natriuretic peptide responses to acute and chronic ventricular pacing in sheep. Am J Physiol. 1996;270:H594–602.PubMed
18.
Zurück zum Zitat Bini GC, Paperini L, Pauletti M, Pardini E, Manfredini E, Corbucci G. A method to calculate tissue impedance through a standard bipolar pacing lead. Cardiovasc Eng. 2006;6:45–52.CrossRef Bini GC, Paperini L, Pauletti M, Pardini E, Manfredini E, Corbucci G. A method to calculate tissue impedance through a standard bipolar pacing lead. Cardiovasc Eng. 2006;6:45–52.CrossRef
19.
Zurück zum Zitat Charles CJ, Nicholls MG, Rademaker MT, Richards AM. Adrenomedullin modulates the neurohumoral response to acute volume loading in normal conscious sheep. J Endocrinol. 2002;173:123–9.PubMedCrossRef Charles CJ, Nicholls MG, Rademaker MT, Richards AM. Adrenomedullin modulates the neurohumoral response to acute volume loading in normal conscious sheep. J Endocrinol. 2002;173:123–9.PubMedCrossRef
20.
Zurück zum Zitat Charles CJ, Tang F, Cameron VA, Richards AM, Espiner EA. Intracerebroventricular atrial natriuretic factor (ANF) antiserum inhibits volume-induced ANF in sheep: evidence for the brain’s regulation of ANF secretion. Endocrinology. 1991;129:2225–30.PubMedCrossRef Charles CJ, Tang F, Cameron VA, Richards AM, Espiner EA. Intracerebroventricular atrial natriuretic factor (ANF) antiserum inhibits volume-induced ANF in sheep: evidence for the brain’s regulation of ANF secretion. Endocrinology. 1991;129:2225–30.PubMedCrossRef
21.
Zurück zum Zitat Nakashima T, Takasugi N, Kubota T, et al. ‘False-positive’ intrathoracic impedance monitor alarm caused by amiodarone-induced hypothyroidism in a patient with cardiac resynchronization therapy-defibrillator. Europace. 2012;14:768–9.PubMedCrossRef Nakashima T, Takasugi N, Kubota T, et al. ‘False-positive’ intrathoracic impedance monitor alarm caused by amiodarone-induced hypothyroidism in a patient with cardiac resynchronization therapy-defibrillator. Europace. 2012;14:768–9.PubMedCrossRef
Metadaten
Titel
Thoracic impedance measures tissue characteristics in the vicinity of the electrodes, not intervening lung water: implications for heart failure monitoring
verfasst von
Christopher J. Charles
Miriam T. Rademaker
Iain C. Melton
Dan Gutfinger
Neal L. Eigler
Fujian Qu
Richard W. Troughton
Publikationsdatum
01.02.2015
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 1/2015
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-014-9570-x

Weitere Artikel der Ausgabe 1/2015

Journal of Clinical Monitoring and Computing 1/2015 Zur Ausgabe

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.