Accuracy of bedside glucometry in critically ill children with peripheral hypoperfusion

verfasst von: Anant Khositseth, Onsuthi Pharadornuwat, Rojjanee Lertbunrian, Nattachai Anantasit, Jarin Vaewpanich

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 1/2015

Einloggen, um Zugang zu erhalten

Abstract

The accuracy of glucose test strip in critically care has been questioned. We investigated the accuracy of glucose test strip in critically ill children. Patients, aged from 1 month to 18 years admitted in pediatric intensive care unit. Demographic data, hemodynamic parameters, and perfusion index (PI) were recorded. Glucose test strips were performed from finger stick blood [capillary blood glucose (CBG)] and from whole blood [whole blood glucose (WBG)] along with laboratory plasma blood glucose (PBG) from either arterial or venous blood samples. The accuracy of glucose test strips was defined according to ISO 15197 and Clarke error grid (CEG). One hundred and eighty one blood samplings including 117 arterial blood (CBG, WBGa, PBGa) and 64 venous blood (CBG, WBGv, PBGv) were obtained. The accuracy of WBGa was 98.3 and 95.2 % when compared to the accuracy of CBG (88.7 and 83.3 %. The accuracy of WBGv was 92.2 % and 87.0 when compared to the accuracy of CBG which was 79.7 and 72.9 % (ISO 15197: 2003 and 2013, respectively). Bland–Altman plot demonstrated bias and precision of 7.4 ± 17.7 mg/dL in acceptable PI group compared to 30.2 ± 23.4 mg/dL in low PI group (PI ≤ 0.3). The CBG test strip must be interpreted carefully in critically ill children. A low PI was associated with poor CBG strip accuracy. WBG test strip from arterial blood was more appropriate for glucose monitoring in children with peripheral hypoperfusion.
Literatur
1.
Zurück zum Zitat Mebis L, Gunst J, Langouche L, Vanhorebeek I, Van den Berghe G. Indication and practical use of intensive insulin therapy in the critically ill. Curr Opin Crit Care. 2007;13:392–8.PubMedCrossRef Mebis L, Gunst J, Langouche L, Vanhorebeek I, Van den Berghe G. Indication and practical use of intensive insulin therapy in the critically ill. Curr Opin Crit Care. 2007;13:392–8.PubMedCrossRef
2.
Zurück zum Zitat Critchell CD, Savarese V, Callahan A, Aboud C, Jabbour S, Marik P. Accuracy of bedside capillary blood glucose measurements in critically ill patients. Intensive Care Med. 2007;33:2079–84.PubMedCrossRef Critchell CD, Savarese V, Callahan A, Aboud C, Jabbour S, Marik P. Accuracy of bedside capillary blood glucose measurements in critically ill patients. Intensive Care Med. 2007;33:2079–84.PubMedCrossRef
3.
Zurück zum Zitat Cengiz E, Tamborlane WV. A tale of two compartments: interstitial versus blood glucose monitoring. Diabetes Technol Ther. 2009;11:S11–6.PubMedCrossRef Cengiz E, Tamborlane WV. A tale of two compartments: interstitial versus blood glucose monitoring. Diabetes Technol Ther. 2009;11:S11–6.PubMedCrossRef
4.
Zurück zum Zitat Karon BS, Gandhi GY, Nuttall GA, Bryant SC, Schaff HV, McMahon MM, et al. Accuracy of roche accu-chek inform whole blood capillary, arterial, and venous glucose values in patients receiving intensive intravenous insulin therapy after cardiac surgery. Am J Clin Pathol. 2007;127:919–26.PubMedCrossRef Karon BS, Gandhi GY, Nuttall GA, Bryant SC, Schaff HV, McMahon MM, et al. Accuracy of roche accu-chek inform whole blood capillary, arterial, and venous glucose values in patients receiving intensive intravenous insulin therapy after cardiac surgery. Am J Clin Pathol. 2007;127:919–26.PubMedCrossRef
5.
Zurück zum Zitat Desachy A, Vuagnat AC, Ghazali AD, Baudin OT, Longuet OH, Calvat SN, et al. Accuracy of bedside glucometry in critically ill patients: influence of clinical characteristics and perfusion index. Mayo Clin Proc. 2008;83:400–5.PubMedCrossRef Desachy A, Vuagnat AC, Ghazali AD, Baudin OT, Longuet OH, Calvat SN, et al. Accuracy of bedside glucometry in critically ill patients: influence of clinical characteristics and perfusion index. Mayo Clin Proc. 2008;83:400–5.PubMedCrossRef
6.
Zurück zum Zitat De Felice C, Goldstein MR, Parrini S, Verrotti A, Criscuolo M, Latini G. Early dynamic changes in pulse oximetry signals in preterm newborns with histologic chorioamnionitis. Pediatr Crit Care Med. 2006;7:138–42.PubMedCrossRef De Felice C, Goldstein MR, Parrini S, Verrotti A, Criscuolo M, Latini G. Early dynamic changes in pulse oximetry signals in preterm newborns with histologic chorioamnionitis. Pediatr Crit Care Med. 2006;7:138–42.PubMedCrossRef
7.
Zurück zum Zitat Lima AP, Beelen P, Bakker J. Use of a peripheral perfusion index derived from the pulse oximetry signal as a noninvasive indicator of perfusion. Crit Care Med. 2002;30:1210–3.PubMedCrossRef Lima AP, Beelen P, Bakker J. Use of a peripheral perfusion index derived from the pulse oximetry signal as a noninvasive indicator of perfusion. Crit Care Med. 2002;30:1210–3.PubMedCrossRef
8.
Zurück zum Zitat Bland JM, Altman DG. Statistical method for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.PubMedCrossRef Bland JM, Altman DG. Statistical method for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.PubMedCrossRef
9.
Zurück zum Zitat ISO 15197. In vitro diagnostic test systems—requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. Geneva, Switzerland: International Organization for Standardization; 2003. ISO 15197. In vitro diagnostic test systems—requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. Geneva, Switzerland: International Organization for Standardization; 2003.
10.
Zurück zum Zitat ISO 15197. In vitro diagnostic test systems—requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. Geneva, Switzerland: International Organization for Standardization; 2013. ISO 15197. In vitro diagnostic test systems—requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. Geneva, Switzerland: International Organization for Standardization; 2013.
11.
Zurück zum Zitat Clark WL, Cox D, Gonder-Frederick LA, Carter WL, Pohl SL. Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care. 1987;10:662–8.CrossRef Clark WL, Cox D, Gonder-Frederick LA, Carter WL, Pohl SL. Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care. 1987;10:662–8.CrossRef
12.
Zurück zum Zitat Colagiuri S, Sandbaek A, Carstensen B, Christensen J, Glumer C, Lauritzen T, et al. Comparability of venous and capillary glucose measurements in blood. Diabet Med. 2003;20:953–6.PubMedCrossRef Colagiuri S, Sandbaek A, Carstensen B, Christensen J, Glumer C, Lauritzen T, et al. Comparability of venous and capillary glucose measurements in blood. Diabet Med. 2003;20:953–6.PubMedCrossRef
13.
Zurück zum Zitat Meex C, Poncin J, Chapelle JP, Cavalier E. Analytical validation of the new plasma calibrated accu-chek test strips (roche diagnostics). Clin Chem Lab Med. 2006;44:1376–8.PubMedCrossRef Meex C, Poncin J, Chapelle JP, Cavalier E. Analytical validation of the new plasma calibrated accu-chek test strips (roche diagnostics). Clin Chem Lab Med. 2006;44:1376–8.PubMedCrossRef
14.
Zurück zum Zitat Meynaar IA, van Spreuwel M, Tangkau PL, Dawson L, Sleeswijk Visser S, Rijks L, et al. Accuracy of accuchek glucose measurement in intensive care patients. Crit Care Med. 2009;37:2691–6.PubMedCrossRef Meynaar IA, van Spreuwel M, Tangkau PL, Dawson L, Sleeswijk Visser S, Rijks L, et al. Accuracy of accuchek glucose measurement in intensive care patients. Crit Care Med. 2009;37:2691–6.PubMedCrossRef
15.
Zurück zum Zitat Kanji S, Buffie J, Hutton B, Bunting PS, Singh A, McDonald K, et al. Reliability of point-of-care testing for glucose measurement in critically ill adults. Crit Care Med. 2005;33:2778–85.PubMedCrossRef Kanji S, Buffie J, Hutton B, Bunting PS, Singh A, McDonald K, et al. Reliability of point-of-care testing for glucose measurement in critically ill adults. Crit Care Med. 2005;33:2778–85.PubMedCrossRef
Metadaten
Titel
Accuracy of bedside glucometry in critically ill children with peripheral hypoperfusion
verfasst von
Anant Khositseth
Onsuthi Pharadornuwat
Rojjanee Lertbunrian
Nattachai Anantasit
Jarin Vaewpanich
Publikationsdatum
01.02.2015
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 1/2015
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-014-9579-1

Weitere Artikel der Ausgabe 1/2015

Journal of Clinical Monitoring and Computing 1/2015 Zur Ausgabe

Bei schweren Reaktionen auf Insektenstiche empfiehlt sich eine spezifische Immuntherapie

Insektenstiche sind bei Erwachsenen die häufigsten Auslöser einer Anaphylaxie. Einen wirksamen Schutz vor schweren anaphylaktischen Reaktionen bietet die allergenspezifische Immuntherapie. Jedoch kommt sie noch viel zu selten zum Einsatz.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.