Erschienen in: Intensive Care Medicine 3/2012

01.03.2012 | Physiological and Technical Notes

In-line filter included into the syringe infusion pump assembly reduces flow irregularities

verfasst von: B. Brotschi, B. Grass, M. Weiss, C. Doell, V. Bernet

Erschienen in: Intensive Care Medicine | Ausgabe 3/2012

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To evaluate whether an in-line filter inserted in the syringe pump infusion line assembly influences start-up times and flow irregularities during vertical pump displacement at low infusion rates.

Methods

Fluid delivery after syringe pump start-up and after vertical displacement of the syringe pump by −50 cm was determined gravimetrically at flow rates of 0.5, 1.0 and 2.0 ml h−1. Measurements were repeated for each flow rate four times with two different syringe pumps with and without an in-line filter incorporated. Data are shown as median and range.

Results

Start-up times were reduced by an in-line filter at 0.5 ml h−1 flow rate from 355.5 s (0–660) to 115 s (0–320), whereas the effect was attenuated at higher flow rates. Pooling of fluid into the infusion system after lowering the infusion syringe pump was halved in all flow rates tested. Amount of infusion bolus after elevating the syringe pump by 50 cm was not affected by an in-line filter.

Conclusion

In the evaluated model in-line filters help to reduce flow irregularities and delay in drug delivery of syringe pumps at low flow rates and represent an option to optimize continuous administration of highly concentrated short-acting drugs at very small infusion rates.
Literatur
1.
Zurück zum Zitat Rooke GA, Bowdle TA (1994) Syringe pumps for infusion of vasoactive drugs: mechanical idiosyncrasies and recommended operating procedures. Anesth Analg 78:150–156PubMedCrossRef Rooke GA, Bowdle TA (1994) Syringe pumps for infusion of vasoactive drugs: mechanical idiosyncrasies and recommended operating procedures. Anesth Analg 78:150–156PubMedCrossRef
2.
Zurück zum Zitat Lonnqvist PA, Lofqvist B (1997) Design flaw can convert commercially available continuous syringe pumps to intermittent bolus injectors. Intensive Care Med 23:998–1001PubMedCrossRef Lonnqvist PA, Lofqvist B (1997) Design flaw can convert commercially available continuous syringe pumps to intermittent bolus injectors. Intensive Care Med 23:998–1001PubMedCrossRef
3.
Zurück zum Zitat Kern H, Kuring A, Redlich U, Dopfmer UR, Sims NM, Spies CD, Kox WJ (2001) Downward movement of syringe pumps reduces syringe output. Br J Anaesth 86:828–831PubMedCrossRef Kern H, Kuring A, Redlich U, Dopfmer UR, Sims NM, Spies CD, Kox WJ (2001) Downward movement of syringe pumps reduces syringe output. Br J Anaesth 86:828–831PubMedCrossRef
4.
Zurück zum Zitat Neff T, Fischer J, Fehr S, Baenziger O, Weiss M (2001) Evaluation of the FASTSTART mode for reducing start-up delay in syringe pump infusion systems. Swiss Med Wkly 131:219–222PubMed Neff T, Fischer J, Fehr S, Baenziger O, Weiss M (2001) Evaluation of the FASTSTART mode for reducing start-up delay in syringe pump infusion systems. Swiss Med Wkly 131:219–222PubMed
5.
Zurück zum Zitat Neff T, Fischer J, Fehr S, Baenziger O, Weiss M (2001) Start-up delays of infusion syringe pumps. Paediatr Anaesth 11:561–565PubMedCrossRef Neff T, Fischer J, Fehr S, Baenziger O, Weiss M (2001) Start-up delays of infusion syringe pumps. Paediatr Anaesth 11:561–565PubMedCrossRef
6.
Zurück zum Zitat Weiss M, Fischer J, Neff T, Baenziger O (2000) The effects of syringe plunger design on drug delivery during vertical displacement of syringe pumps. Anaesthesia 55:1094–1098PubMedCrossRef Weiss M, Fischer J, Neff T, Baenziger O (2000) The effects of syringe plunger design on drug delivery during vertical displacement of syringe pumps. Anaesthesia 55:1094–1098PubMedCrossRef
7.
Zurück zum Zitat Neff TA, Fischer JE, Schulz G, Baenziger O, Weiss M (2001) Infusion pump performance with vertical displacement: effect of syringe pump and assembly type. Intensive Care Med 27:287–291PubMedCrossRef Neff TA, Fischer JE, Schulz G, Baenziger O, Weiss M (2001) Infusion pump performance with vertical displacement: effect of syringe pump and assembly type. Intensive Care Med 27:287–291PubMedCrossRef
8.
Zurück zum Zitat Weiss M, Baenziger O, Neff T, Fanconi S (2000) Influence of infusion line compliance on drug delivery rate during acute line loop formation. Intensive Care Med 26:776–779PubMedCrossRef Weiss M, Baenziger O, Neff T, Fanconi S (2000) Influence of infusion line compliance on drug delivery rate during acute line loop formation. Intensive Care Med 26:776–779PubMedCrossRef
9.
Zurück zum Zitat Weiss M, Hug MI, Neff T, Fischer J (2000) Syringe size and flow rate affect drug delivery from syringe pumps. Can J Anesth 47:1031–1035PubMedCrossRef Weiss M, Hug MI, Neff T, Fischer J (2000) Syringe size and flow rate affect drug delivery from syringe pumps. Can J Anesth 47:1031–1035PubMedCrossRef
10.
Zurück zum Zitat Cook RI (2000) Syringe pump assemblies and the natural history of clinical technology. Can J Anesth 47:929–935PubMedCrossRef Cook RI (2000) Syringe pump assemblies and the natural history of clinical technology. Can J Anesth 47:929–935PubMedCrossRef
11.
Zurück zum Zitat Leff RD, True WR, Roberts RJ (1987) A gravimetric technique for evaluating flow continuity from two infusion devices. Am J Hosp Pharm 44:1388–1391PubMed Leff RD, True WR, Roberts RJ (1987) A gravimetric technique for evaluating flow continuity from two infusion devices. Am J Hosp Pharm 44:1388–1391PubMed
12.
Zurück zum Zitat Trautmann M, Zauser B, Wiedeck H, Buutenschön K, Marre R (2001) Bacterial colonization and endotoxin contamination of intravenous fluids. J Hosp Infect 37:225–236CrossRef Trautmann M, Zauser B, Wiedeck H, Buutenschön K, Marre R (2001) Bacterial colonization and endotoxin contamination of intravenous fluids. J Hosp Infect 37:225–236CrossRef
13.
Zurück zum Zitat Jack T, Brent BE, Boehne M, Müller M, Sewald K, Braun A, Wessel A, Sasse M (2010) Analysis of particulate contaminations of infusion solutions in a pediatric intensive care unit. Intensive Care Med 36:707–711PubMedCrossRef Jack T, Brent BE, Boehne M, Müller M, Sewald K, Braun A, Wessel A, Sasse M (2010) Analysis of particulate contaminations of infusion solutions in a pediatric intensive care unit. Intensive Care Med 36:707–711PubMedCrossRef
14.
Zurück zum Zitat Kunac DL, Ball PA, Broadbet RS (1999) In-line intravenous filtration in neonates: help not hindrance. Aust J Hosp Pharm 29:321–327 Kunac DL, Ball PA, Broadbet RS (1999) In-line intravenous filtration in neonates: help not hindrance. Aust J Hosp Pharm 29:321–327
15.
Zurück zum Zitat Ball PA (2003) Intravenous in-line filters: filtering the evidence. Curr Opin Clin Nutr Metab Care 6:319–325PubMed Ball PA (2003) Intravenous in-line filters: filtering the evidence. Curr Opin Clin Nutr Metab Care 6:319–325PubMed
Metadaten
Titel
In-line filter included into the syringe infusion pump assembly reduces flow irregularities
verfasst von
B. Brotschi
B. Grass
M. Weiss
C. Doell
V. Bernet
Publikationsdatum
01.03.2012
Verlag
Springer-Verlag
Erschienen in
Intensive Care Medicine / Ausgabe 3/2012
Print ISSN: 0342-4642
Elektronische ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-011-2452-5

Weitere Artikel der Ausgabe 3/2012

Intensive Care Medicine 3/2012 Zur Ausgabe

Delir bei kritisch Kranken – Antipsychotika versus Placebo

16.05.2024 Delir Nachrichten

Um die Langzeitfolgen eines Delirs bei kritisch Kranken zu mildern, wird vielerorts auf eine Akuttherapie mit Antipsychotika gesetzt. Eine US-amerikanische Forschungsgruppe äußert jetzt erhebliche Vorbehalte gegen dieses Vorgehen. Denn es gibt neue Daten zum Langzeiteffekt von Haloperidol bzw. Ziprasidon versus Placebo.

Eingreifen von Umstehenden rettet vor Erstickungstod

15.05.2024 Fremdkörperaspiration Nachrichten

Wer sich an einem Essensrest verschluckt und um Luft ringt, benötigt vor allem rasche Hilfe. Dass Umstehende nur in jedem zweiten Erstickungsnotfall bereit waren, diese zu leisten, ist das ernüchternde Ergebnis einer Beobachtungsstudie aus Japan. Doch es gibt auch eine gute Nachricht.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.