Navigated laparoscopic microwave ablation of tumour mimics in pig livers: a randomized ex-vivo experimental trial

verfasst von: M. N. Thomas, G. Dieplinger, R. R. Datta, R. Kleinert, H. F. Fuchs, A. Bunck, M. Peterhans, C. J. Bruns, D. Stippel, R. Wahba

Erschienen in: Surgical Endoscopy | Ausgabe 12/2021

Abstract

Background

In order to efficiently perform laparoscopic microwave ablation of liver tumours precise positioning of the ablation probe is mandatory. This study evaluates the precision and ablation accuracy using the innovative laparoscopic stereotactic navigation system CAS-One-SPOT in comparison to 2d ultrasound guided laparoscopic ablation procedures.

Methods

In a pig liver ablation model four surgeons, experienced (n = 2) and inexperienced (n = 2) in laparoscopic ablation procedures, were randomized for 2d ultrasound guided laparoscopic or stereotactic navigated laparoscopic ablation procedures. Each surgeon performed a total of 20 ablations. Total attempts of needle placements, time from tumor localization till beginning of ablation and ablation accuracy were analyzed.

Results

The use of the laparoscopic stereotactic navigation system led to a significant reduction in total attempts of needle placement. The experienced group of surgeons reduced the mean number of attempts from 2.75 ± 2.291 in the 2d ultrasound guided ablation group to 1.45 ± 1.191 (p = 0.0302) attempts in the stereotactic navigation group. Comparable results could be observed in the inexperienced group with a reduction of 2.5 ± 1.50 to 1.15 ± 0.489 (p = 0.0005). This was accompanied by a significant time saving from 101.3 ± 112.1 s to 48.75 ± 27.76 s (p = 0.0491) in the experienced and 165.5 ± 98.9 s to 66.75 ± 21.96 s (p < 0.0001) in the inexperienced surgeon group. The accuracy of the ablation process was hereby not impaired as postinterventional sectioning of the ablation zone revealed.

Conclusion

The use of a stereotactic navigation system for laparoscopic microwave ablation procedures of liver tumors significantly reduces the attempts and time of predicted correct needle placement for novices and experienced surgeons without impairing the accuracy of the ablation procedure.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Radiofrequency and microwave ablations are increasingly used to treat primary and secondary liver malignancies as an alternative to surgery or as an adjunct to surgical resection in patients with impaired liver function, technically non-resectable liver tumours, distinct bilobar tumour manifestation and patients in which large liver resections cannot be safely performed due to concurrent medical conditions [1].
Currently the microwave ablation applicator is placed in a laparoscopic or open procedure under 2D ultrasound guidance into the lesion.
Accurate positioning of the applicator into the lesion is crucial in order (a) to achieve complete ablation of the targeted lesion and (b) to minimize the risk of injury of adjacent structures. This technique is associated with a significant learning curve and explains the reported high local recurrence rates [2].
Tumour targeting and ablation applicator placement is an even greater problem particularly in laparoscopic ablation procedures, since the intuitive spatial correlation of the hand holding the ultrasound probe and the hand placing the applicator in the tumour is displayed on two different screens [3].
In addition, major complications following liver thermoablation including intra-peritoneal bleeding, portal or hepatic vein thrombosis, bile duct injury, liver dysfunction, and intestinal perforation remain a clinically relevant problem in up to 10% of patients [4, 5].
These complications are often directly linked to the trauma associated with applicator placement. Even oncologic outcome could be deteriorated by inadequate applicator placement, as the risk of tumour seeding in the needle tract is linked directly to multiple needle insertions [6]. Therefore, the number of attempts for proper applicator placement should be minimized during the ablation procedure and is of utmost importance. The use of 2D-ultrasound for proper applicator placement in a 3D-space requires a high degree of expertise and skills. There is a significant learning curve in thermoablation of liver tumours. After 50 performed RFA procedures for liver tumours complication rate drops and the rate of complete ablation increases [7]. Especially highly experienced hepatopancreatobiliary surgeons were able to hit a target in a single pass using 2D-ultrasound guidance in as low as 59% of the cases [8].
It could be demonstrated that the use of magnetic and infrared 3D guided ablation techniques led to significantly higher accuracy in applicator placement during simulated ablation procedures [810].
The aim of this study was to evaluate targeting precision and ablation accuracy using the innovative laparoscopic stereotactic navigation system CAS-One-SPOT in comparison to 2d ultrasound guided laparoscopic ablation in a randomized experimental ex vivo porcine tumour mimic model (ClinicalTrials.gov: NCT04106453).

Methods

SPOT software

The CAS-One-SPOT system (CAScination AG, Bern, Switzerland) is an innovative electromagnetic tracking and navigation system for laparoscopic microwave ablation. The device is comprised of a navigation platform which connects to a table-mounted, electromagnetic field generator. The position of a 4-way articulated, laparoscopic ultrasound probe (BK Medical, Herlev, Denmark) is tracked by a dedicated adapter clip including an electromagnetic sensor. In parallel, the position of an ablation needle is tracked using a sensor attached to the needle shaft by means of an adapter clip. The navigation software then calculates the expected intersection point between the needle path and the ultrasound image and displays the position of the needle tip as well as the expected ablation volume. In the first step, the user is asked to define a target point for the ablation needle on the ultrasound image (Fig. 1, left). In the second step, needle guidance towards the target point is provided by means of a bullseye targeting display and a live ultrasound image with position of a simulated ablation zone is displayed (Fig. 1, right). The software documents the target points selected and the positioning of the needle/ablation zone relative to the target (Fig. 1). With the SPOT system, targeting and ablation needle placement is possible from any angle. In contrast, using the conventional laparoscopic technique, the needle placement must be performed in-plane of the ultrasound device.

Tumour mimics

Tumour mimics were generated in pig livers as described earlier by Scott et al. [11]. Briefly, a gelantine-like mixture of 3% agarose, 3% cellulose, 7% glycerol, and 0.05% methylene blue was prepared. Before injection, the mixture was heated to 70 °C and a constant volume was injected into the pre-cooled (4 °C) pig livers forming intrahepatic tumour lesions of 1–1.5 cm size. No more than 6 tumour mimics were injected into each pig liver. The tumour mimics have been earlier evaluated for radiofrequency ablation in our working group [10].

Study design

After tumour mimic injection the pig livers (weighing about 1500 g) were fixed on a cork tile and placed into a laparoscopic torso trainer. Ports for the laparoscopic camera (Einstein Vision, B.Braun, Germany) and the laparoscopic ultrasound probe (BK Medical, USA) were placed through a foam plate on the top front of the laparoscopic trainer, simulating the rigidity of the abdominal wall (Fig. 2).
This IRB-approved study was designed as a randomized cross-over trial.
A total of 4 surgeons performed the experiments. They were separated into 2 groups. Group 1 surgeons (n = 2) were novice to the ablation technique and group 2 surgeons (n = 2) were experienced (over 200 and 1000 ablations in human). Each surgeon performed a total of 20 ablations. After initial ultrasound localisation of the hepatic lesion the ablation process was randomized in 2D ultrasound guided or 3D navigated using the SPOT software. Time was measured from first puncture of the foam plate until beginning of ablation for each targeted lesion. Additionally, the number of attempts for correct needle placement was documented. Ablations were performed using the SOLERO (Angiodynamics, USA) microwave tissue ablation system. In order to keep the ablation zone small enough to see differences in the accuracy of needle placement in the different groups, an ablation power of 100 W for 2 min was chosen. After ablation the localisation of the targeted lesion was visualized on the liver surface in order to differentiate between the multiple intrahepatic lesions. For further analyses the pig livers were taken out of the torso trainer and each ablation zone was cut in an standardized axial and transverse plane as described by Mulier et al. [7]. Then the specimens were placed on millimetre paper, photographed and digitalized for further measurements (Figs. 3, 4).
Measurements were performed using Image J software v 1.52 following the recommendation of Mulier et al. [7] of a standardized description of coagulation size. Furthermore, the investigator was blinded to the ablation process that was used—SPOT vs. 2d ultrasound guided. In the axial plane the axial diameter, the maximal transverse diameter, the front margin and the Euclidian distance (centre of the ablation zone to centre of the tumour lesion) were measured (Fig. 5). In the transverse plane the maximal transverse diameters and the Euclidian distance were measured. Measurements were performed with the open source software Image J v 1.52. The study was registered at clininicaltrial.gov (NCT04106453).

Statistical analyses

Statistical analyses were performed using IBM SPSS 23. Graph Pad Prism 6 was used to create figures. Data distributions were assumed to be Gaussian and summarised as mean ± standard deviation and compared by two-sided (Student’s) t-tests for paired or for unpaired data.

Results

Number of targeting attempts

The mean number of targeting attempts until predicted correct positioning of the needle by the inexperienced surgeons using the standard ultrasound 2D guided technique was 2.5 ± 1.5 attempts. With the use of the stereotactic navigation system the mean number of attempts could be reduced to 1.15 ± 0.48 attempts (p = 0.0005). The experienced group of surgeons needed a mean of 2.75 ± 2.29 attempts in the conventional ablation setting and were able to reduce the number of attempts to a mean of 1.45 ± 1.19 in the stereotactic navigated ablation group (Fig. 6) (p = 0.0302).

Time until ablation

The inexperienced surgeons needed a mean time of 165.5 ± 98.9 s in the 2d ultrasound guided ablation group from first introduction of the needle until predicted correct positioning of the needle. By using the stereotactic navigation system this could be significantly reduced to a mean of 66.75 ± 21.96 s (p < 0.0001).
In line, the experienced surgeons were able to reduce the mean time from first needle puncture to beginning of ablation from a mean of 101.3 ± 112.1 s in the 2d guided ultrasound group to a mean of 48.75 ± 27.76 s in the stereotactic navigated group (p = 0.0491) (Fig. 6).

Targeting accuracy

Volumetric measurements of the ablation zones revealed no statistically significant difference between the groups (2d ultrasound guided inexperienced vs. stereotactic guidance inexperienced 4.9 ± 0.52cm2 vs. 4.33 ± 0.41cm2/2d ultrasound guided experienced vs. stereotactic guidance experienced 5.27 ± 0.68cm2 vs. 7.17 ± 0.83cm2). Additionally the front margin (end of tumour to end of ablation) (2d ultrasound guided inexperienced vs. stereotactic guidance inexperienced 0.09 cm ± 0.69 vs. 0.22 ± 0.61 (p = 0.543)/2d ultrasound guided experienced vs. stereotactic guidance experienced − 0.10 cm ± 0.51 vs. − 0.06 cm ± 0.50 (p = 0.838)) nor the Eucledian-distance (2d ultrasound guided inexperienced vs. stereotactic guidance inexperienced 0.95 cm ± 0.46 vs. 0.89 ± 0.62 (p = 0.73)/2d ultrasound guided experienced vs. stereotactic guidance experienced 1.06 cm ± 0.41 vs. 0.92 cm ± 0.30 (p = 0.228)) revealed no statistically significant measurements between the analysed groups.

Discussion

Fast and accurate placement of the ablation electrode is crucial in laparoscopic microwave ablation for malignant liver tumours to reduce surgical complications and provide optimal oncologic outcome. Using the conventional 2D ultrasound guided technique a relevant learning curve has to be passed through. Targeting systems seem to help surgeons here. In general, using the system is complex. Our study demonstrates that the use of an electromagnetic 3D guided ultrasounds navigation system in laparoscopic microwave ablation procedures leads to a significant reduction in attempts until predicted correct needle placement is accomplished. In our experimental setting the use of the 3D guided ultrasound ablation method led to a significant reduction in needle withdrawals of 49.5% attempts. This reduction was accompanied by a significant time saving of 43% from localisation of the intrahepatic target to the beginning of ablation. Even the group of experienced surgeons had high benefit from the SPOT system compared to the conventional ultrasound technique. Needle placement becomes fast and easy. Furthermore, the targeting accuracy did not significantly differ. Although experienced surgeon were faster in the placement of the applicator, interestingly both groups of surgeons, experienced and inexperienced surgeons, had comparable results regarding total attempts of needle placement and accuracy of the ablation procedure even in the conventional ultrasound guided ablation setting. This can in part be explained by the used model of porcine livers. Even if this model is widely accepted for the studying and learning of thermal ablation techniques [10, 11], porcine livers are relatively flat and tumour mimics are predominantly injected near to the liver surface. Therefore, difficult angels and long trajectories through healthy liver parenchyma in order to hit the lesion are rare. The experimental use of bovine livers could increase the degree of difficulty and possibly reveal more differences between the two groups.
Thermal ablation of liver tumours has evolved to be a broadly accepted tool in the management of increasingly complex oncologic patients. However, this procedure is still afflicted with the occurrence of major complications [4, 10].
One feared complication of open, laparoscopic and percutaneous ablation procedures is the release of neoplastic cells along the needle track. This complication occurs in up to 2.5% [6, 12, 13]. The risk of this complication is directly linked to the need of multiple needle placements until correct positioning of the needle is achieved in the centre of the tumour [6, 14]. With the use of the 3d navigated ablation procedure this complication could be possibly diminished due to the significantly reduced number of needle insertions.
Krasnick et al. [9] could demonstrate comparable results in an in vivo porcine model of 3D electromagnetic guided ultrasound ablations. In the used porcine model, a fiducial marker was placed in the liver and kidney and ablation was performed either using the 3D guided ultrasound technique or the conventional 2D guided ultrasound technique. Ablation procedures were conducted by 3 general surgeons experienced in ablation procedures (> 48 ablations performed in clinical routine). In their experimental laparoscopic setup, the use of the electromagnetic guided 3D ultrasound ablation technology led to a significant reduction of 62% in needle placement attempts until the ablation procedure could be conducted as compared to the conventional 2D guided ultrasound ablation. The targeting accuracy, as defined by locating the fiducial marker in the ablation zone, was not significantly different in the analysed groups. The specimen analyses in order to study the accuracy of the performed ablation procedure was carried out in a single plane. However, as the ablation zone is a 3D-elipsoid structure a second sectional plane (transversal and axial) should have been analysed in order to reliably study targeting accuracy. In our experimental setting the tumour mimic localisation was analysed in relation to the ablation zone in 2 sectional planes. We could demonstrate that the use of the 3D guided electromagnetic field in laparoscopic liver ablation was not associated with a diminished targeting accuracy.
Paolucci et al. analysed an experimental electromagnetic ultrasound-based navigation technique for experimental laparoscopic ablation of liver tumor mimics. They could show with an experimental electromagnetic ultrasound-based navigation technique a significant reduction in time for positioning of the ablation probe and a significant reduction in needed repositioning attempts of the ablation needle [15]. However no real ablation of a tumour was performed in this study: only the adequate positioning of the needle was calculated based on a 3D ultrasound image.
Another recently published study by Tinguely et al. analysed the clinical use of laparoscopic image-guided microwave ablation technique for malignant liver tumours [16]. In their retrospective study ablation procedures were performed using an intraoperative rigid four-landmark-based registration technique from a pre-operatively reconstructed 3d model of the liver. This technique however is accompanied with a high degree of technical and financial effort as 3d model reconstruction of the liver is mandatory for every patient before the planed ablation procedure.
The laparoscopic use of the CASOne-Spot System in liver ablations procedures allows precise targeting of liver lesions and is linked to a significant reduction in necessary needle insertions. With the use of this new technology complications observed in conventional 2D guided laparoscopic ablation procedures can be potentially minimized and wider indications for laparoscopic ablation can be achieved.

Conclusion

The use of an electromagnetic-navigated 3D ultrasound ablation technique in laparoscopic ablation procedures, when compared to the 2D ultrasound guidance, leads to a significant reduction in necessary attempts of needle placement without restricting ablation accuracy. With the clinical use of this new technique laparoscopic microwave ablation could be simplified, thus wider indications for laparoscopic ablations can be developed and complications associated with needle placement could be diminished.

Compliance with ethical standards

Disclosures

Drs. Michael Thomas, G. Dieplinger, R.R. Datta, R. Kleinert, H.F. Fuchs, A. Bunck,, C.J. Bruns, D.Stippel, R. Wahba have no conflicts of interest or financial ties to disclose. Dr. M. Peterhans reports personal fees from CAScination AG and was involved in the development of the hardware and software of the CAS-One-SPOT system.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

e.Med Interdisziplinär

Kombi-Abonnement

Jetzt e.Med zum Sonderpreis bestellen!

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Jetzt bestellen und 100 € sparen!

Literatur
1.
Zurück zum Zitat Martin RCG, Scoggins CR, McMasters KM (2010) Safety and efficacy of microwave ablation of hepatic tumors: a prospective review of a 5-year experience. Ann Surg Oncol 17:171–178CrossRef Martin RCG, Scoggins CR, McMasters KM (2010) Safety and efficacy of microwave ablation of hepatic tumors: a prospective review of a 5-year experience. Ann Surg Oncol 17:171–178CrossRef
3.
Zurück zum Zitat Santambrogio R, Barabino M, D’Alessandro V, Galfrascoli E, Zappa MA, Piccolo G, Zuin M, Opocher E (2020) Laparoscopic thermoablation for hepatocellular carcinoma in patients with liver cirrhosis: an effective procedure for tricky tumors. Med Oncol 37:1–7CrossRef Santambrogio R, Barabino M, D’Alessandro V, Galfrascoli E, Zappa MA, Piccolo G, Zuin M, Opocher E (2020) Laparoscopic thermoablation for hepatocellular carcinoma in patients with liver cirrhosis: an effective procedure for tricky tumors. Med Oncol 37:1–7CrossRef
7.
Zurück zum Zitat Mulier S, Ni Y, Frich L, Burdio F, Denys AL, de Wispelaere JF, Dupas B, Habib N, Hoey M, Jansen MC, Lacrosse M, Leveillee R, Miao Y, Mulier P, Mutter D, Ng KK, Santambrogio R, Stippel D, Tamaki K, van Gulik TM, Marchal G, Michel L (2007) Experimental and clinical radiofrequency ablation: proposal for standardized description of coagulation size and geometry. Ann Surg Oncol 14:1381–1396. https://doi.org/10.1245/s10434-006-9033-9CrossRefPubMed Mulier S, Ni Y, Frich L, Burdio F, Denys AL, de Wispelaere JF, Dupas B, Habib N, Hoey M, Jansen MC, Lacrosse M, Leveillee R, Miao Y, Mulier P, Mutter D, Ng KK, Santambrogio R, Stippel D, Tamaki K, van Gulik TM, Marchal G, Michel L (2007) Experimental and clinical radiofrequency ablation: proposal for standardized description of coagulation size and geometry. Ann Surg Oncol 14:1381–1396. https://​doi.​org/​10.​1245/​s10434-006-9033-9CrossRefPubMed
Metadaten
Titel
Navigated laparoscopic microwave ablation of tumour mimics in pig livers: a randomized ex-vivo experimental trial
verfasst von
M. N. Thomas
G. Dieplinger
R. R. Datta
R. Kleinert
H. F. Fuchs
A. Bunck
M. Peterhans
C. J. Bruns
D. Stippel
R. Wahba
Publikationsdatum
07.12.2020
Verlag
Springer US
Erschienen in
Surgical Endoscopy / Ausgabe 12/2021
Print ISSN: 0930-2794
Elektronische ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-020-08180-5

Weitere Artikel der Ausgabe 12/2021

Surgical Endoscopy 12/2021 Zur Ausgabe

Häusliche Gewalt in der orthopädischen Notaufnahme oft nicht erkannt

28.05.2024 Häusliche Gewalt Nachrichten

In der Notaufnahme wird die Chance, Opfer von häuslicher Gewalt zu identifizieren, von Orthopäden und Orthopädinnen offenbar zu wenig genutzt. Darauf deuten die Ergebnisse einer Fragebogenstudie an der Sahlgrenska-Universität in Schweden hin.

Fehlerkultur in der Medizin – Offenheit zählt!

28.05.2024 Fehlerkultur Podcast

Darüber reden und aus Fehlern lernen, sollte das Motto in der Medizin lauten. Und zwar nicht nur im Sinne der Patientensicherheit. Eine negative Fehlerkultur kann auch die Behandelnden ernsthaft krank machen, warnt Prof. Dr. Reinhard Strametz. Ein Plädoyer und ein Leitfaden für den offenen Umgang mit kritischen Ereignissen in Medizin und Pflege.

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

TAVI versus Klappenchirurgie: Neue Vergleichsstudie sorgt für Erstaunen

21.05.2024 TAVI Nachrichten

Bei schwerer Aortenstenose und obstruktiver KHK empfehlen die Leitlinien derzeit eine chirurgische Kombi-Behandlung aus Klappenersatz plus Bypass-OP. Diese Empfehlung wird allerdings jetzt durch eine aktuelle Studie infrage gestellt – mit überraschender Deutlichkeit.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.