To the Editor:
Intraoperative, salvaged, autologous blood re-transfusions are especially useful to treat profuse hemorrhage that occurs during obstetric surgeries [1]. A prominent concern with such autotransfusions is the risk that the relatively large pore size (170–260 μm) of standard blood transfusion filters may not eliminate all amniotic and fetal materials and other contaminants from the salvaged, autologous blood where they can cause amniotic fluid embolization (AFE).
Use of leukocyte reduction filters (LRFs) with 10–50 μm pore size during salvage of autologous blood in cesarean sections appears to be able to eliminate amniotic and fetal particulate substances such as fetal squamous cells [2], and thereby prevent AFE. Intraoperative autotransfusions with LRFs, therefore, represent a promising technique for the treatment of profuse hemorrhage in obstetric surgeries. However, since 2007, the Japanese Red Cross Blood Center has provided pre-storage leukocyte-reduced red cell concentrate containing a citrate–phosphate–dextrose solution instead of an acid–citrate–dextrose solution as anticoagulant in which the leukocytes were reduced soon after collection. LRFs have not been made since 2006, and have not been supplied since 2009 by Japanese manufacturers.
Microaggregate blood filters (MBFs) with 10–40 μm pore size are able to eliminate larger-diameter, retained, blood component microaggregate materials and non-blood, particulate matter that is potentially harmful to the recipients. In other words, MBFs assure the patient protection from microaggregate materials, clots, and particulate debris that can cause pulmonary embolization [3].
We managed a case with a re-transfusion of salvaged, autologous blood using a Pall SQ40® (Pall, Tokyo, Japan) MBF with 40 μm pore size instead of LRF for massive hemorrhage during obstetric surgery [4]. Finally, abortion and total abdominal hysterectomy were performed and no AFE or other complications occurred.
The Pall SQ40® MBF is a free-flow, low priming volume device which provides a 40 μm rated filter and media pleating for the elimination of potentially harmful blood component microaggregates and non-blood component particulate matter [5]. Pall’s 40 μm rated polyester screen media is process-controlled and monitored to ensure uniformity of pore size to assure maximum patient protection from microaggregates, clots, and particulate debris.
Further, screen-type MBFs are able to maintain a relatively high speed of transfusion, because of their larger total filter area. The total filter area of Pall’s screen-type MBF (160 cm2) provides up to twelve times the typical first surface area of a depth-type MBF. This feature reduces the number of administration set changes required for rapid transfusion, particularly for blood components with a large debris load such as salvaged blood components. In massive obstetric hemorrhage, rapid re-transfusions are required for resuscitation. Since LRFs limit the speed of transfusion, it is sometimes necessary to suspend or forego the use of an LRF, and to use screen-type MBFs, instead, in order to increase the speed of transfusion.
In conclusion, MBFs instead of LRFs should be available and ready to use to purify salvaged, autologous blood for re-transfusion to treat profuse hemorrhage during obstetric surgeries, and to prevent AFE.

Conflict of interest

There are no financial interests or conflicts of interest to declare.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.
download
DOWNLOAD
print
DRUCKEN

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Anästhesiologie

Kombi-Abonnement

Mit e.Med Anästhesiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes AINS, den Premium-Inhalten der AINS-Fachzeitschriften, inklusive einer gedruckten AINS-Zeitschrift Ihrer Wahl.

Literatur
1.
2.
Zurück zum Zitat Waters JH, Biscotti C, Potter PS, Phillipson E. Amniotic fluid removal during cell salvage in the cesarean section patient. Anesthesiology. 2000;92:1531–6.PubMedCrossRef Waters JH, Biscotti C, Potter PS, Phillipson E. Amniotic fluid removal during cell salvage in the cesarean section patient. Anesthesiology. 2000;92:1531–6.PubMedCrossRef
3.
Zurück zum Zitat Reul GJ Jr, Beall AC Jr, Greenberg SD. Protection of the pulmonary microvasculature by fine screen blood filtration. Chest. 1974;66:4–9.PubMedCrossRef Reul GJ Jr, Beall AC Jr, Greenberg SD. Protection of the pulmonary microvasculature by fine screen blood filtration. Chest. 1974;66:4–9.PubMedCrossRef
4.
Zurück zum Zitat Morita Y, Mizuno J, Takada S, Yunokawa S, Morita S. Massive hemorrhage during abdominal total hysterectomy in a patient with placenta percreta. Masui. 2009;58:1025–7.PubMed Morita Y, Mizuno J, Takada S, Yunokawa S, Morita S. Massive hemorrhage during abdominal total hysterectomy in a patient with placenta percreta. Masui. 2009;58:1025–7.PubMed
5.
Zurück zum Zitat Cullen DJ, Kunsman J, Caldera D, Dennis RC, Valeri CR. Comparative evaluation of new fine-screen filters: effects on blood flow rate and microaggregate removal. Anesthesiology. 1980;53:3–8.PubMedCrossRef Cullen DJ, Kunsman J, Caldera D, Dennis RC, Valeri CR. Comparative evaluation of new fine-screen filters: effects on blood flow rate and microaggregate removal. Anesthesiology. 1980;53:3–8.PubMedCrossRef
Metadaten
Titel
Use of microaggregate blood filters instead of leukocyte reduction filters to purify salvaged, autologous blood for re-transfusion during obstetric surgery
verfasst von
Ju Mizuno
Publikationsdatum
01.08.2013
Verlag
Springer Japan
Erschienen in
Journal of Anesthesia / Ausgabe 4/2013
Print ISSN: 0913-8668
Elektronische ISSN: 1438-8359
DOI
https://doi.org/10.1007/s00540-013-1579-7

Weitere Artikel der Ausgabe 4/2013

Journal of Anesthesia 4/2013 Zur Ausgabe

Ähnliche Überlebensraten nach Reanimation während des Transports bzw. vor Ort

29.05.2024 Reanimation im Kindesalter Nachrichten

Laut einer Studie aus den USA und Kanada scheint es bei der Reanimation von Kindern außerhalb einer Klinik keinen Unterschied für das Überleben zu machen, ob die Wiederbelebungsmaßnahmen während des Transports in die Klinik stattfinden oder vor Ort ausgeführt werden. Jedoch gibt es dabei einige Einschränkungen und eine wichtige Ausnahme.

Häusliche Gewalt in der orthopädischen Notaufnahme oft nicht erkannt

28.05.2024 Häusliche Gewalt Nachrichten

In der Notaufnahme wird die Chance, Opfer von häuslicher Gewalt zu identifizieren, von Orthopäden und Orthopädinnen offenbar zu wenig genutzt. Darauf deuten die Ergebnisse einer Fragebogenstudie an der Sahlgrenska-Universität in Schweden hin.

Fehlerkultur in der Medizin – Offenheit zählt!

28.05.2024 Fehlerkultur Podcast

Darüber reden und aus Fehlern lernen, sollte das Motto in der Medizin lauten. Und zwar nicht nur im Sinne der Patientensicherheit. Eine negative Fehlerkultur kann auch die Behandelnden ernsthaft krank machen, warnt Prof. Dr. Reinhard Strametz. Ein Plädoyer und ein Leitfaden für den offenen Umgang mit kritischen Ereignissen in Medizin und Pflege.

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.