Detection of spine structures with Bioimpedance Probe (BIP) Needle in clinical lumbar punctures

verfasst von: Sanna Halonen, Kari Annala, Juho Kari, Samuli Jokinen, Aki Lumme, Kai Kronström, Arvi Yli-Hankala

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 5/2017

Einloggen, um Zugang zu erhalten

Abstract

Lumbar puncture is a relatively safe procedure, but some serious, even fatal, complications can occur. Needle guidance can increase puncture accuracy, decrease the number of attempts, and make the procedure easier. We tested the feasibility of a bioimpedance-based tissue-sensing technology for needle guidance in clinical use. The Bioimpedance Probe (BIP) Needle has a removable BIP stylet enabling measurement of bioimpedance spectra during the procedure. The BIP Needle is connected to a measurement device that uses tissue-classification software, and the device provides audiovisual feedback when it detects cerebrospinal fluid (CSF). We performed spinal anesthesia with the BIP Needle in 45 patients. The device performance and needle tip location were verified by an experienced anesthesiologist confirming CSF leakage. The device detected CSF in all cases (sensitivity of 100 %). Six cases with false detections lowered the specificity to 81 %, but in practice, most of these were easy to differentiate from true detections because their duration was short and they occurred during backward movement of the needle. The epidural spectrum differentiated as fatty tissue from surrounding tissues, but the ligamentum flavum was not clearly detectable in the data. The BIP Needle is a reliable tool for detecting CSF in lumbar puncture. It can make the puncture procedure smoother, as repeated CSF flow tests are avoided. The correct needle tip location is immediately detected, thus unnecessary needle movements close to spinal nerves are prevented. Physicians could benefit from the information provided by the BIP Needle, especially in patients with obesity or anatomic alterations.
Literatur
1.
Zurück zum Zitat Hermanides J, Hollmann MW, Stevens MF, Lirk P. Failed epidural: causes and management. Br J Anaesth. 2012;109(2):144–54.CrossRefPubMed Hermanides J, Hollmann MW, Stevens MF, Lirk P. Failed epidural: causes and management. Br J Anaesth. 2012;109(2):144–54.CrossRefPubMed
2.
Zurück zum Zitat Hyderally H. Complications of spinal anesthesia. Mt Sinai J Med. 2001;69(1–2):55–6. Hyderally H. Complications of spinal anesthesia. Mt Sinai J Med. 2001;69(1–2):55–6.
3.
Zurück zum Zitat Kang XH, Bao FP, Xiong XX, Li M, Jin TT, Shao J, Zhu SM. Major complications of epidural anesthesia: a prospective study of 5083 cases at a single hospital. Acta Anaesth Scand. 2014;58(7):858–66.CrossRefPubMed Kang XH, Bao FP, Xiong XX, Li M, Jin TT, Shao J, Zhu SM. Major complications of epidural anesthesia: a prospective study of 5083 cases at a single hospital. Acta Anaesth Scand. 2014;58(7):858–66.CrossRefPubMed
4.
Zurück zum Zitat Pitkänen MT, Aromaa U, Cozanitis DA, Förster JG. Serious complications associated with spinal and epidural anaesthesia in Finland from 2000 to 2009. Acta Anaesth Scand. 2013;57(5):553–64.CrossRefPubMed Pitkänen MT, Aromaa U, Cozanitis DA, Förster JG. Serious complications associated with spinal and epidural anaesthesia in Finland from 2000 to 2009. Acta Anaesth Scand. 2013;57(5):553–64.CrossRefPubMed
5.
Zurück zum Zitat Chin KJ, Perlas A, Chan V, Brown-Shreves D, Koshkin A, Vaishnav V. Ultrasound imaging facilitates spinal anesthesia in adults with difficult surface anatomic landmarks. Anesthesiology. 2011;115(1):94–101.CrossRefPubMed Chin KJ, Perlas A, Chan V, Brown-Shreves D, Koshkin A, Vaishnav V. Ultrasound imaging facilitates spinal anesthesia in adults with difficult surface anatomic landmarks. Anesthesiology. 2011;115(1):94–101.CrossRefPubMed
6.
Zurück zum Zitat Seeberger MD, Kaufmann M, Staender S, Schneider M, Scheidegger D. Repeated dural punctures increase the incidence of postdural puncture headache. Anesth Analg. 1996;82(2):302–5.PubMed Seeberger MD, Kaufmann M, Staender S, Schneider M, Scheidegger D. Repeated dural punctures increase the incidence of postdural puncture headache. Anesth Analg. 1996;82(2):302–5.PubMed
7.
Zurück zum Zitat Kopacz DJ, Neal JM, Pollock JE. The regional anesthesia “learning curve”. What is the minimum number of epidural and spinal blocks to reach consistency? Reg Anesth. 1995;21(3):182–90. Kopacz DJ, Neal JM, Pollock JE. The regional anesthesia “learning curve”. What is the minimum number of epidural and spinal blocks to reach consistency? Reg Anesth. 1995;21(3):182–90.
8.
Zurück zum Zitat Kalvøy H, Frich L, Grimnes S, Martinsen ØG, Hol PK, Stubhaug A. Impedance-based tissue discrimination for needle guidance. Physiol Meas. 2009;30(2):129–40.CrossRefPubMed Kalvøy H, Frich L, Grimnes S, Martinsen ØG, Hol PK, Stubhaug A. Impedance-based tissue discrimination for needle guidance. Physiol Meas. 2009;30(2):129–40.CrossRefPubMed
9.
Zurück zum Zitat Trebbels D, Fellhauer F, Jugl M, Haimerl G, Min M, Zengerle R. Online tissue discrimination for transcutaneous needle guidance applications using broadband impedance spectroscopy. IEEE Trans Biomed Eng. 2012;59(2):494–503.CrossRefPubMed Trebbels D, Fellhauer F, Jugl M, Haimerl G, Min M, Zengerle R. Online tissue discrimination for transcutaneous needle guidance applications using broadband impedance spectroscopy. IEEE Trans Biomed Eng. 2012;59(2):494–503.CrossRefPubMed
10.
Zurück zum Zitat Kalvøy H, Sauter AR. Detection of intraneural needle-placement with multiple frequency bioimpedance monitoring: a novel method. J Clin Monitor Comput. 2015;. doi:10.1007/s10877-015-9698-3. Kalvøy H, Sauter AR. Detection of intraneural needle-placement with multiple frequency bioimpedance monitoring: a novel method. J Clin Monitor Comput. 2015;. doi:10.​1007/​s10877-015-9698-3.
11.
Zurück zum Zitat Mishra V, Schned AR, Hartov A, Heaney JA, Seigne J, Halter RJ. Electrical property sensing biopsy needle for prostate cancer detection. Prostate. 2013;73(15):1603–13.PubMed Mishra V, Schned AR, Hartov A, Heaney JA, Seigne J, Halter RJ. Electrical property sensing biopsy needle for prostate cancer detection. Prostate. 2013;73(15):1603–13.PubMed
12.
Zurück zum Zitat Hernandez DJ, Sinkov VA, Roberts WW, Allaf ME, Patriciu A, Jarrett TW, et al. Measurement of bio-impedance with a smart needle to confirm percutaneous kidney access. J Urol. 2001;166(4):1520–3.CrossRefPubMed Hernandez DJ, Sinkov VA, Roberts WW, Allaf ME, Patriciu A, Jarrett TW, et al. Measurement of bio-impedance with a smart needle to confirm percutaneous kidney access. J Urol. 2001;166(4):1520–3.CrossRefPubMed
14.
Zurück zum Zitat Grimness S, Martinsen ØG. Bioimpedance and bioelectricity basics. 2nd ed. London: Academic Press; 2008. Grimness S, Martinsen ØG. Bioimpedance and bioelectricity basics. 2nd ed. London: Academic Press; 2008.
15.
Zurück zum Zitat Ragheb T, Geddes LA. The polarization impedance of common electrode metals operated at low current density. Ann Biomed Eng. 1991;19(2):151–63.CrossRefPubMed Ragheb T, Geddes LA. The polarization impedance of common electrode metals operated at low current density. Ann Biomed Eng. 1991;19(2):151–63.CrossRefPubMed
16.
Zurück zum Zitat Schwan HP. Linear and nonlinear electrode polarization and biological materials. Ann Biomed Eng. 1992;20(3):269–88.CrossRefPubMed Schwan HP. Linear and nonlinear electrode polarization and biological materials. Ann Biomed Eng. 1992;20(3):269–88.CrossRefPubMed
17.
Zurück zum Zitat Gabriel S, Lau RW, Gabriel C. The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz to 20 GHz. Phys Med Biol. 1996;41(11):2251–69.CrossRefPubMed Gabriel S, Lau RW, Gabriel C. The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz to 20 GHz. Phys Med Biol. 1996;41(11):2251–69.CrossRefPubMed
18.
Zurück zum Zitat Mirtaheri P, Grimnes S, Martinsen ØG. Electrode polarization impedance in weak NaCl aqueous solutions. IEEE Trans Biomed Eng. 2005;52(12):2093–9.CrossRefPubMed Mirtaheri P, Grimnes S, Martinsen ØG. Electrode polarization impedance in weak NaCl aqueous solutions. IEEE Trans Biomed Eng. 2005;52(12):2093–9.CrossRefPubMed
19.
Zurück zum Zitat Duda RO, Hart PE, Stork DG. Pattern classification. 2nd ed. New York: John Wiley & Sons; 2012. Duda RO, Hart PE, Stork DG. Pattern classification. 2nd ed. New York: John Wiley & Sons; 2012.
20.
Zurück zum Zitat De Andrés J, Reina MA, Prats A. Epidural space and regional anesthesia. Eur J Pain Suppl. 2009;3(2):55–63.CrossRef De Andrés J, Reina MA, Prats A. Epidural space and regional anesthesia. Eur J Pain Suppl. 2009;3(2):55–63.CrossRef
21.
Zurück zum Zitat Baumann SB, Wozny DR, Kelly SK, Meno FM. The electrical conductivity of human cerebrospinal fluid at body temperature. IEEE Trans Biomed Eng. 1997;44(3):220–3.CrossRefPubMed Baumann SB, Wozny DR, Kelly SK, Meno FM. The electrical conductivity of human cerebrospinal fluid at body temperature. IEEE Trans Biomed Eng. 1997;44(3):220–3.CrossRefPubMed
22.
Zurück zum Zitat Gabriel C, Gabriel S, Corthout E. The dielectric properties of biological tissues: I. Literature survey. Phys Med Biol. 1996;41(11):2231–349.CrossRefPubMed Gabriel C, Gabriel S, Corthout E. The dielectric properties of biological tissues: I. Literature survey. Phys Med Biol. 1996;41(11):2231–349.CrossRefPubMed
Metadaten
Titel
Detection of spine structures with Bioimpedance Probe (BIP) Needle in clinical lumbar punctures
verfasst von
Sanna Halonen
Kari Annala
Juho Kari
Samuli Jokinen
Aki Lumme
Kai Kronström
Arvi Yli-Hankala
Publikationsdatum
04.08.2016
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 5/2017
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-016-9915-8

Weitere Artikel der Ausgabe 5/2017

Journal of Clinical Monitoring and Computing 5/2017 Zur Ausgabe

Blutdrucksenkung schon im Rettungswagen bei akutem Schlaganfall?

31.05.2024 Apoplex Nachrichten

Der optimale Ansatz für die Blutdruckkontrolle bei Patientinnen und Patienten mit akutem Schlaganfall ist noch nicht gefunden. Ob sich eine frühzeitige Therapie der Hypertonie noch während des Transports in die Klinik lohnt, hat jetzt eine Studie aus China untersucht.

Ähnliche Überlebensraten nach Reanimation während des Transports bzw. vor Ort

29.05.2024 Reanimation im Kindesalter Nachrichten

Laut einer Studie aus den USA und Kanada scheint es bei der Reanimation von Kindern außerhalb einer Klinik keinen Unterschied für das Überleben zu machen, ob die Wiederbelebungsmaßnahmen während des Transports in die Klinik stattfinden oder vor Ort ausgeführt werden. Jedoch gibt es dabei einige Einschränkungen und eine wichtige Ausnahme.

Nicht Creutzfeldt Jakob, sondern Abführtee-Vergiftung

29.05.2024 Hyponatriämie Nachrichten

Eine ältere Frau trinkt regelmäßig Sennesblättertee gegen ihre Verstopfung. Der scheint plötzlich gut zu wirken. Auf Durchfall und Erbrechen folgt allerdings eine Hyponatriämie. Nach deren Korrektur kommt es plötzlich zu progredienten Kognitions- und Verhaltensstörungen.

Häusliche Gewalt in der orthopädischen Notaufnahme oft nicht erkannt

28.05.2024 Häusliche Gewalt Nachrichten

In der Notaufnahme wird die Chance, Opfer von häuslicher Gewalt zu identifizieren, von Orthopäden und Orthopädinnen offenbar zu wenig genutzt. Darauf deuten die Ergebnisse einer Fragebogenstudie an der Sahlgrenska-Universität in Schweden hin.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.